Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
FirmwareVersion | 4022 | Firmware Version | Field Device Software Revision- This revision corresponds to the software or firmware, that is embedded in the Field Device | R | ASCII | 2052 |
ElapsedTime | 406A | Elapsed Time | Elapsed time since most recent significant event. | R | Integer | 2049 |
EventLog_faultflags | 406B | Event status | Bit encoded event status stored in log. | R | Enum | 2049 |
EventLog_evt_interval | 406C | Event Interval | Interval associated with a particular event that is the time elapsed since the previous event. | R | Integer | 2049 |
volume_units | 4076 | Volume Units | Select units of measurement for volume output. | R/W | Enum | 8 |
volume_value | 4088 | Vol | Volume of material in vessel | R | Float | 34 |
maximum_volume | 4073 | Maximum Volume | Select maximum volume for application. | R/W | Enum | 8 |
StrapTblLngth | 4075 | Strap Table Length | Number of valid level-volume pairs in strapping table. A level-volume pair is valid if both values are greater than the values of the preceding level-volume pair. | R | Integer | 2048 |
LvlPt01 | 4077 | Pt01Lvl | R/W | Float | 2048 | |
LvlPt02 | 4078 | Pt02Lvl | R/W | Float | 2048 | |
Name | ID | Label | Description | Type | Access | Class |