transfer_function 1F84 转换功能 枚举 32
lvl_set4mA 404E 4mA Set Point 下限值——定义模拟值和范围0%点开始的操作端点。此外,下限值还定义与模拟量值关联的报警和与数值表示关联的报警开始的操作端点。 读/写 浮点 64
lvl_set20mA 404F 20mA Set Point 上限值——定义模拟值和范围100%点开始的操作端点。此外,上限值还定义与模拟量值关联的报警和与数值表示关联的报警开始的操作端点。 读/写 浮点 64
sensor_serial_number 4005 传感器序列号 无符号 2048
ProbeLength 400E Probe Length Distance from sensor reference point to end of probe. 读/写 浮点 2048
LevelOffset 4050 Level Offset For top-mounted sensor, Level Offset is the distance from the zero level point to the end of probe (negative if zero level point is above the end of probe; positive if below.) For bottom-mounted sensor, Level Offset is the distance from the sensor reference point to the zero level point (positive if zero level point is above the reference; negative if below.) 读/写 浮点 2048
SensorMount 401C Sensor Mount Mounting configuration of sensor - selects reference used to determine level. 读/写 枚举 2048
TrimLevel 4051 Trim Level Trim Level is a parameter for fine adjustment of the Level output to a known value. This should be used only after all other parameters have been entered correctly, and it has been confirmed that the transmitter is tracking the level correctly. Gross adjustment should be made using Level Offset. 读/写 浮点 2048
TrimIfcLvl 4052 Trim Ifc Level Trim Interface Level is a parameter for fine adjustment of the IfcLvl output to a known value. This should be used only after all other parameters have been entered correctly, and it has been confirmed that the transmitter is tracking the interface level correctly. Gross adjustments should be made using Level Offset. 读/写 浮点 2048
MeasType 4053 Measurement Type Single float operation (Level Only or Interface Only) reports level of float nearest sensor. Dual float operation (Level and Ifc or Ifc and Level) requires two floats 读/写 枚举 2048