Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
sensor_units | 4000 | Range unit | Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit | R/W | Enum | 8 |
sensor_damping_value | 400B | Damp | Damping Value- Damping time constant applied to both the Digital Current Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. | R/W | Float | 8 |
analog_value | 4041 | AO abs. | Analog Output Value- Value that tracks the Digital Value representation, under normal operating modes | R/W | Float | 66 |
input_percent_range | 4003 | AO rel. | Percent of Range- Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. | R | Float | 34 |
input_upper_value | 4043 | PV URV | Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. | R/W | Float | 32 |
input_lower_value | 4044 | PV LRV | Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. | R/W | Float | 32 |
temp_value | 4045 | S-temp. | Temp.- Temperature measured by the device | R | Float | 10 |
temp_units | 401D | Temp unit | Unit - The temperature unit | R/W | Enum | 8 |
sensor_upper_limit | 4016 | PV USL | Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. | R | Float | 8 |
sensor_lower_limit | 4015 | PV LSL | Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. | R | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |