Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
pressure_lower_range_value | 1397 | Pres LRV | Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. Pressure- Pressure of the process measured with respect to a reference pressure. | R/W | Float | 32 |
pressure_damping_value | 1398 | Pres damp | Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. Pressure- Pressure of the process measured with respect to a reference pressure. | R/W | Float | 8 |
flange_type | 139A | Flange type | Flange Type- Hardware, adjacent to the sensor, that physically connects the process to the sensor. Pressure- Pressure of the process measured with respect to a reference pressure. | R/W | Enum | 2048 |
o_ring_material | 139B | O ring matl | O Ring Material- Construction material of a seal that exists between the module and flange. | R/W | Enum | 2048 |
meter_installation | 139C | Meter type | Meter Installation- Defines whether there is a meter installed or not and possibly what information the meter will display. | R/W | Enum | 256 |
remote_seal_type | 139D | RS type | Remote Seal Type- Type of device that is capable of detecting and passing on the process pressure on to the module. This is only unambiguous if 1 remote seal is installed. | R/W | Enum | 2048 |
module_fill_fluid | 139E | Fill fluid | Sensor Fill Fluid- Fluid material in the module. | R | Enum | 2048 |
number_of_remote_seals | 139F | Num remote seal | Number of Remote Seals- Physical quantity of the seals for the remote seal that is installed. | R/W | Enum | 2048 |
response_preambles | 0FE2 | Num resp preams | Number of Response Preambles- Number of Preambles to be sent in the response message from the Field Device to the Host | W | Unsigned | 128 |
xmtr_specific_status_0 | 0FB5 | Status group 1 | Field Device Specific Status Byte 1- Data that has Field Device status information relating to: failures / warnings / status of process's. | R | BitEnum | 2051 |
Name | ID | Label | Description | Type | Access | Class |