Loading…
Name ID Label Description Type Access Class
compat_burst_trig_class_2 4113 Classification The classification of the burst trigger variable R/W Enum 16
compat_burst_trig_units_2 4114 Unit The unit of the burst trigger variable R/W Enum 16
loc_OptionRangeOutputCurrentA 4356 Current Out. Range Select the output range (Fct. C4.1.4) R/W Enum 4096
bkp_OptionRangeOutputCurrentA 436E Current Out. Range Select the output range (Fct. C4.1.4) R/W Enum 4096
loc_upperRange_value 4354 PV URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 4096
bkp_upperRange_value 436D PV URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 4096
loc_OptionSignalAlarmOutputCurrentA 4357 Error Function Specify error current. (Fct. C4.1.5) R/W Enum 4096
bkp_OptionSignalAlarmOutputCurrentA 436F Error Function Specify error current. (Fct. C4.1.5) R/W Enum 4096
loc_lowerRange_value 4355 PV LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 4096
bkp_lowerRange_value 4371 PV LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 4096
Name ID Label Description Type Access Class