Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
varOffsetCustomer_Value | 41B8 | Offset | Direct adjustment of the offset for compensation of the mounting position | R/W | Float | 2048 |
varConfigInterfaceMeasuring_DensityUpperMedium | 41B9 | Density upper medium | Requirement for the interface measurement is that both sensors are covered by the medium. The interface can only be detected between the two sensors. | R/W | Float | 2048 |
varConfigInterfaceMeasuring_DensityLowerMedium | 41BB | Density lower medium | Requirement for the interface measurement is that both sensors are covered by the medium. The interface can only be detected between the two sensors. | R/W | Float | 2048 |
varConfigInterfaceMeasuring_DistanceBetweenSensors | 41BC | Distance of the sensors (A) | The distance of the two sensors must be entered for determination of the interface out of the measured differential pressure | R/W | Float | 2048 |
varConfigDensityMeasuring_DistanceBetweenSensors | 41BD | Distance of the sensors (A) | The distance of the two sensors must be entered for determination of the density out of the measured differential pressure | R/W | Float | 2048 |
varDeviceHardwareVersionAscii_VersionInfoAscii | 41BE | Hardware version | Hardware version of the main electronics | R | ASCII | 2048 |
varDeviceSoftwareVersionAscii_VersionInfoAscii | 41BF | Software version | Software version of the main electronics | R | ASCII | 2048 |
varCalibrationDate_CalibrationDay | 41C0 | R | Unsigned | 2048 | ||
varCalibrationDate_CalibrationMonth | 41C2 | R | Unsigned | 2048 | ||
varCalibrationDate_CalibrationYear | 41C3 | R | Unsigned | 2048 | ||
Name | ID | Label | Description | Type | Access | Class |