Loading…
Name ID Label Description Type Access Class
utpMassFlow 4047 Update Time Update Time Period 'UTP'. Indicates the maximum period between updates of the process variable. R Time 8
simulateMassFlow 4048 Set Simulation Desired setting of the process variable simulation. If enabled, the process variable will be simulated and won't show the actual process value. W Enum 2048
propertiesMassFlow 4049 Properties Properties, indicate e.g. the simulation state. R BitEnum 10
dfoTemperature 404D Format Temperature Display format for Temperature. Local for HART applications, no correlation with device's local display. R/W Enum 4096
dvlTemperature 404C Temperature Shows the current measurement for Temperature. R/W Float 10
dmpTemperature 404E Time constant Time constant for Temperature R/W Float 8
uslTemperature 404F Upper Sensor Limit Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
lslTemperature 4050 Lower Sensor Limit Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
mspTemperature 4051 Minimum Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
clsTemperature 4052 Unit Class Temperature Device Variable Class. Defines also the set of selectable physical units. R Enum 8
Name ID Label Description Type Access Class