名称 ID 标签 描述 类型 访问 分类
uniViscosity 40BC Viscosity 枚举 2048
flCompressibilityFactorFluid 40BD Compressibility Compressibility factor; specifies the deviation between the real gas and ideal gas. 浮点 2048
enuLimitSwitch 40C3 读/写 索引 4096
enuB1Function 40C7 Actual function (B1) Switchpoint actual function; specifies the allocation for the binary output B1 (inactive / limitswitch primary measurement / limitswitch totalizer / pulse output*) (* The pulse output is only available in case of an IO module). 读/写 枚举 32
flB1FlowLimit 40C9 PV switch-point (B1) PV switch point limit; range 0.0 ... upper range value, if current primary measurement exceeds the appointed switchpoint, the output changes the binary status. 读/写 浮点 32
flB1FlowHysteresis 40CB PV switch-point hysteresis (B1) PV switch point hysteresis; range 0.0 ... switchpoint, if current primary measurement exceeds the appointed switchpoint, the output changes the binary status. To return to initial position, the switchpoint value reduced by the hysteresis must be under-run. 读/写 浮点 32
flB1TotalizerLimit 40CE Totalizer switch-point (B1) SV switchpoint limit; range 0.0 ... totalizer limit, if the current totalizer value exceeds adjusted limit switch value, the output changes the binary status. There is no hysteresis adjustment for this switchpoint. 读/写 浮点 32
enuB1ContactType 40D1 Type of contact (B1) Switchpoint type of contact; the binary limit switch output can be assigned to normally closed contact or normally open contact. 读/写 枚举 2048
enuB2Function 40D3 Actual function (B2) Switchpoint actual function; specifies the allocation for the binary output B2 (inactive / limitswitch primary measurement / limitswitch totalizer / pulse output*) (* The pulse output is only available in case of an IO module). 读/写 枚举 32
flB2FlowLimit 40D4 PV switch-point (B2) PV switch point limit; range 0.0 ... upper range value, if current primary measurement exceeds the appointed switchpoint, the output changes the binary status. 读/写 浮点 32
名称 ID 标签 描述 类型 访问 分类