Loading…
Name ID Label Description Type Access Class
dunVolFlow 400D Volume Flow unit Primary measurement unit; physical unit for the measurand of the active process (e.g. volume flow) as well as for the corresponding limit switch - INFORMATION: changing of this parameter has no effect on the linearization area. R/W Enum 32
dunNormalizedVolFlow 4020 Volume Flow unit Primary measurement unit; physical unit for the measurand of the active process (e.g. volume flow) as well as for the corresponding limit switch - INFORMATION: changing of this parameter has no effect on the linearization area. R/W Enum 32
dunMassFlow 402F Mass Flow unit Primary measurement unit; physical unit for the measurand of the active process (e.g. volume flow) as well as for the corresponding limit switch - INFORMATION: changing of this parameter has no effect on the linearization area. R/W Enum 32
dunLevel 403E Level measurement unit Primary measurement unit; physical unit for the measurand of the active process (e.g. volume flow) as well as for the corresponding limit switch - INFORMATION: changing of this parameter has no effect on the linearization area. R/W Enum 32
dvlVolFlow 400E Volume Flow PV value; measured value for primary measurand (PV). R/W Float 34
dvlNormalizedVolFlow 4021 Volume Flow PV value; measured value for primary measurand (PV). R/W Float 34
dvlMassFlow 4030 Mass Flow PV value; measured value for primary measurand (PV). R/W Float 34
dvlLevel 403F Level PV value; measured value for primary measurand (PV). R/W Float 34
dfoVolFlow 400F Format Defines format specifier for measurements and configuration. Parameters having measurements units: "Full Scale", ranges and sensor limits. Affects only HART applications (FC375/FC475 screens, AMS faceplates/views, PDM, FDM...) but not instrument's local display. R/W Enum 4096
dfoNormalizedVolFlow 4022 Format Defines format specifier for measurements and configuration. Parameters having measurements units: "Full Scale", ranges and sensor limits. Affects only HART applications (FC375/FC475 screens, AMS faceplates/views, PDM, FDM...) but not instrument's local display. R/W Enum 4096
Name ID Label Description Type Access Class