Name ID Label Description Type Access Class
sensor_serial_number 400A PV Snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R Unsigned 2048
level_value 400D Level Level - The total product measurement of the device R Float 10
level_units 400E This is the UNIT of measure for the process variable of the device and provides a basis for all of the setup functions. Selectable engineering UNITS include: inches, feet, meters, millimeters and centimeters. R/W Enum 8
UnitsNone 400F Units R Enum 2048
lin_level_value 4010 Linearized Level Linearized Level - The output of the linearization table based on Level R Float 10
interface_value 42AC Interface Interface - The measured value of the interface product R Float 10
lin_interface_value 42AD Linearized Interface Linearized Interface - The output of the linearization table based on Interface R Float 10
floodedchambermode 42AE Flooded Chamber Mode A flooded condition means the entire probe is submerged in the process. A non-flooded condition means a portion of the probe above the process is covered bay a gas layer. R/W Enum 2048
fixedtotallevel 42AF Fixed Total Level Flooded total replaces the total level measurement in the calculation of the interface level measurement. R/W Float 2048
sw_revision_date 4019 Software Revision Date Software Revision Date- Gregorian calendar date denoting the date of Software Release R/W HartDate 2048