Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
B_val | 4000 | B Stern-Geary | Derived from the Stern-Geary Relationship. If set to 0, the default of 0.0256 is used. Only used in output calculation if Meth2 is selected. | R/W | Float | 0 |
K_val | 4001 | K Comb Props | Combined material and corrosion reaction properties. See the manual to determine this value. | R/W | Float | 0 |
A_val | 4002 | A Elec Area | The area of one electrode in square centimeters. | R/W | Float | 0 |
F_Freq | 4003 | Filter Freq | Line frequency for ADC low pass filter. | R/W | Enum | 0 |
PF_Serial | 4004 | Serial Number | P+F Device Serial Number | R/W | ASCII | 0 |
Preference | 4005 | Preference | Select the variable to control the output. 0 = Type1, 1 = Type2 or Localization. | R/W | Enum | 0 |
PV_Alarm | 4006 | Alarm Config | Select the alarm type. Issued when output exceeds URV or LRV. Will never be issued if the output is Type2. Manual alarm is reset when the Write Alarm, Command 145, is sent. | R/W | Enum | 0 |
PF_PW | 4007 | PF Pswd | Eight character password | W | ASCII | 0 |
IC_PW | 4008 | IC Pswd | Eight character password | W | ASCII | 0 |
Mea_Mode | 4009 | Meas Mode | Corrosion measurement mode. Normally set to 7, Meter Mode. | R/W | Enum | 0 |
Calc_Method | 400A | Method | Calculation method. 0 = Meth1, 1 = Meth2 | R/W | Enum | 0 |
Test_Mode | 400B | Method | Calculation method. 0 = Meth1, 1 = Meth2 | R/W | Enum | 0 |
Comm_Date | 400C | Comm Date | Manufacturing Commission Date | R/W | HartDate | 0 |
Sens_Data0 | 400D | Data [0] | Results from the measurement process. | R | Float | 2051 |
Sens_Data1 | 400E | Data [1] | Results from the measurement process. | R | Float | 2051 |
Sens_Data2 | 400F | Data [2] | Results from the measurement process. | R | Float | 2051 |
Sens_Data3 | 4010 | Data [3] | Results from the measurement process. | R | Float | 2051 |
Sens_Data4 | 4011 | Data [4] | Results from the measurement process. | R | Float | 2051 |
Sens_Data5 | 4012 | Data [5] | Results from the measurement process. | R | Float | 2051 |
Sens_Data6 | 4013 | Data [6] | Results from the measurement process. | R | Float | 2051 |
Sens_Data7 | 4014 | Data [7] | Results from the measurement process. | R | Float | 2051 |
Sens_Data8 | 4015 | Data [8] | Results from the measurement process. | R | Float | 2051 |
Sens_Data9 | 4016 | Data [9] | Results from the measurement process. | R | Float | 2051 |
Sens_DataA | 4017 | Data [A] | Results from the measurement process. | R | Float | 2051 |
Sens_DataB | 4018 | Data [B] | Results from the measurement process. | R | Float | 2051 |
Sens_DataC | 4019 | Data [C] | Results from the measurement process. | R | Float | 2051 |
Sens_DataD | 401A | Data [D] | Results from the measurement process. | R | Float | 2051 |
Mea_OffsetV | 401B | V Offset | Voltage Channel Offset. Typically 625 millivolts +/- 6. | R/W | Float | 0 |
Mea_OffsetI | 401C | I Offset | Current Channel Offset. Typically 625 millivolts +/- 6. | R/W | Float | 0 |
PV_Avg | 401D | PV Avg | Average of the selected measurement over the number of mearurement cycles since these were last cleared. | R | Float | 18 |
PV_Cnt | 401E | Samp Count | The number of measurement cycles since last cleared. | R | Unsigned | 18 |
PV_val | 401F | PV | Selected measurement output. Type1 in mm/yr or mil/yr or Type2 | R | Float | 34 |
PV_urv | 4020 | URV | Selected measurement upper output limit. Not applicable if Preference is set to Type2. If the alarm is enabled, the loop current will go to alarm condition if the measured value exceeds this limit. | R/W | Float | 0 |
Sens_hilim | 4021 | Sens Max | Sensor selected measurement maximum. 1000.0 for mm/yr, 40000.0 for mil/yr, 1.0 for Preference set to Type2 | R | Float | 0 |
PV_lrv | 4022 | LRV | Selected measurement lower output limit. Not applicable if Preference is set to Type2. If the alarm is enabled, the loop current will go to alarm condition if the measured value exceeds this limit. | R/W | Float | 0 |
Sens_lolim | 4023 | Sens Min | Sensor selected measurement minimum. 0.0 for any output selection | R | Float | 0 |
PV_span | 4024 | Min Span | Selected measurement minimum span. 1.0 if mm/yr, 40.0 if mil/yr. Not applicable if Preference is set to Type2. | R | Float | 0 |
PV_mA | 4025 | AO1 | AO (アナログ出力値) - ノーマル オペレーティング モードでのデジタル値で表示します。 | R/W | Float | 66 |
PV_percent | 4026 | PV % レンジ | % レンジ - ノーマル オペレーティング モードでの Lower/Upper レンジ値によるデジタル値表示です。単位は常にパーセントです。 | R | Float | 34 |
PV_damp | 4027 | Damping | Measurement damping time in seconds. | R/W | Float | 8 |
PV_Xfr | 4028 | Xfr Code | Transfer Function Code | R | Enum | 0 |
PV_Units | 4029 | PV Units | Selected Ouput Variable Units if Type1, mm/yr or mil/yr, else if Type2, none | R/W | Enum | 0 |
sensor_serial_number | 402A | PV センサー S/N | センサー シリアル ナンバー - デジタル値表示又はフィールド機器バリアブルを主に導出するセンサーを一意に識別します。 | R | Unsigned | 2048 |
Dummy_password | 402B | Dummy Password | R | ASCII | 4096 | |
Dummy_units | 402C | Dummy Units | R | Unsigned | 2048 | |
MV_PW | 402D | MV Pswd | Eight character password | W | ASCII | 0 |
MV_PWstate | 402E | MV Pswd St | Password enabled indicator, 0 = diasbled, 1 = enabled | R | Enum | 0 |
MV_Mode | 402F | MV Mode | MV Mode. Disabled or Enabled. | W | Enum | 0 |
SV_val | 4031 | Local Corrosion | Secondary measurement output. | R | Float | 34 |
TV_val | 4032 | Stern-Geary | Tertiary measurement output. | R | Float | 34 |
Name | ID | Label | Description | Type | Access | Class |