Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
Sens_DataC | 4019 | Data [C] | Results from the measurement process. | R | Float | 2051 |
Sens_DataD | 401A | Data [D] | Results from the measurement process. | R | Float | 2051 |
Mea_OffsetV | 401B | V Offset | Voltage Channel Offset. Typically 625 millivolts +/- 6. | R/W | Float | 0 |
Mea_OffsetI | 401C | I Offset | Current Channel Offset. Typically 625 millivolts +/- 6. | R/W | Float | 0 |
PV_Avg | 401D | PV Avg | Average of the selected measurement over the number of mearurement cycles since these were last cleared. | R | Float | 18 |
PV_Cnt | 401E | Samp Count | The number of measurement cycles since last cleared. | R | Unsigned | 18 |
PV_val | 401F | PV | Selected measurement output. Type1 in mm/yr or mil/yr or Type2 | R | Float | 34 |
PV_urv | 4020 | URV | Selected measurement upper output limit. Not applicable if Preference is set to Type2. If the alarm is enabled, the loop current will go to alarm condition if the measured value exceeds this limit. | R/W | Float | 0 |
Sens_hilim | 4021 | Sens Max | Sensor selected measurement maximum. 1000.0 for mm/yr, 40000.0 for mil/yr, 1.0 for Preference set to Type2 | R | Float | 0 |
PV_lrv | 4022 | LRV | Selected measurement lower output limit. Not applicable if Preference is set to Type2. If the alarm is enabled, the loop current will go to alarm condition if the measured value exceeds this limit. | R/W | Float | 0 |
Sens_lolim | 4023 | Sens Min | Sensor selected measurement minimum. 0.0 for any output selection | R | Float | 0 |
PV_span | 4024 | Min Span | Selected measurement minimum span. 1.0 if mm/yr, 40.0 if mil/yr. Not applicable if Preference is set to Type2. | R | Float | 0 |
PV_mA | 4025 | AO1 | Valor de salida analógica: Valor que sigue la representación del valor digital, bajo modos de funcionamiento normales | R/W | Float | 66 |
PV_percent | 4026 | PV % rnge | Porcentaje de rango: Variable que rastrea la representación del valor digital con respecto al rango definido por el valor de rango inferior y el valor de rango superior, para modos de funcionamiento normales. Las unidades de esta variable siempre son en porcentaje. | R | Float | 34 |
PV_damp | 4027 | Damping | Measurement damping time in seconds. | R/W | Float | 8 |
PV_Xfr | 4028 | Xfr Code | Transfer Function Code | R | Enum | 0 |
PV_Units | 4029 | PV Units | Selected Ouput Variable Units if Type1, mm/yr or mil/yr, else if Type2, none | R/W | Enum | 0 |
sensor_serial_number | 402A | PV Snsr s/n | Número de serie del sensor: Identifica de manera única el sensor del que se deriva principalmente la representación del valor digital o la variable del dispositivo de campo. | R | Unsigned | 2048 |
Dummy_password | 402B | Dummy Password | R | ASCII | 4096 | |
Dummy_units | 402C | Dummy Units | R | Unsigned | 2048 | |
MV_PW | 402D | MV Pswd | Eight character password | W | ASCII | 0 |
MV_PWstate | 402E | MV Pswd St | Password enabled indicator, 0 = diasbled, 1 = enabled | R | Enum | 0 |
MV_Mode | 402F | MV Mode | MV Mode. Disabled or Enabled. | W | Enum | 0 |
SV_val | 4031 | Local Corrosion | Secondary measurement output. | R | Float | 34 |
TV_val | 4032 | Stern-Geary | Tertiary measurement output. | R | Float | 34 |
Name | ID | Label | Description | Type | Access | Class |