Loading…
名称 ID 标签 描述 类型 访问 分类
reset_password_code 434F Reset Password code Allows to enter reset password code 无符号 4096
master_password 4350 Reset Password code Allows to enter reset password code Password 4096
facta_urv 421D Cal A URV URV during factory A calibration 浮点 8
factb_urv 421E Cal B URV URV during factory B calibration 浮点 8
factc_urv 421F Cal C URV URV during factory C calibration 浮点 8
facta_lrv 4220 Cal A LRV LRV during factory A calibration 浮点 8
factb_lrv 4221 Cal B LRV LRV during factory B calibration 浮点 8
factc_lrv 4222 Cal C LRV LRV during factory C calibration 浮点 8
request_sel_cal 4218 Calibration Selection Allows user to select one of the available Calibration options when CAL A&B or CAL A,B&C are available. When only CAL A option is available in the device, default factory calibration range is active in the device. When CAL B and/or C is available, then the user can select the desired Calibration option: CAL A, B or C. 读/写 枚举 8
actual_sel_cal 4219 Active Calibration Actual Calibration used by the sensor: CAL A, CAL B or CAL C. When only CAL A option is available in the device, default factory calibration range is active in the device. When CAL B and/or C is available, then the user selected Calibration option is the current active calibration. Range is CAL A, CAL B or CAL C. 枚举 8
名称 ID 标签 描述 类型 访问 分类