Name ID Label Description Type Access Class
rSetpointEventB 4135 Setpoint Event B Sets the Event Controller Setpoint for Relay B. R/W Float 0
rSetpointEventC 4136 Setpoint Event C Sets the Event Controller Setpoint for Relay C. R/W Float 0
rSetpointEventD 4137 Setpoint Event D Sets the Event Controller Setpoint for Relay D. R/W Float 0
rDeadbandEventA 4138 Sets the Controller Deadband/Dead Zone for Relay B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal. R/W Float 0
rDeadbandEventB 4139 Sets the Controller Deadband/Dead Zone for Relay B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal R/W Float 0
rDeadbandEventC 413A Sets the Controller Deadband/Dead Zone for Relay C. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal R/W Float 0
rDeadbandEventD 413B Sets the Controller Deadband/Dead Zone for Relay D. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal R/W Float 0
rPIManualPWMA 413C Manual Value A Sets the Manual control output level (%) Relay A. R/W Float 0
rPIManualPWMB 413E Manual Value B Sets the Manual control output level (%) Relay B. R/W Float 0
rPIManualPWMC 4140 Manual Value C Sets the Manual control output level (%) Relay C. R/W Float 0
Name ID Label Description Type Access Class