Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
eRange2 | 40D8 | Range Output 2 | Selection List of sensor ranges. | R | Enum | 0 |
iOneLowFloatIndexSC200 | 40DC | R/W | Index | 4096 | ||
rHighAlarmAlarmA | 40DE | High Alarm A | Sets the High Alarm setpoint for Alarm A. | R/W | Float | 0 |
rHighAlarmAlarmB | 40E1 | High Alarm B | Sets the High Alarm setpoint for Alarm B. | R/W | Float | 0 |
rHighAlarmAlarmC | 40E4 | High Alarm C | Sets the High Alarm setpoint for Alarm C. | R/W | Float | 0 |
rHighAlarmAlarmD | 40E7 | High Alarm D | Sets the High Alarm setpoint for Alarm D. | R/W | Float | 0 |
rLowAlarmAlarmA | 40EA | Low Alarm A | Sets the Low Alarm setpoint for Alarm A. | R/W | Float | 0 |
rLowAlarmAlarmB | 40EB | Low Alarm B | Sets the Low Alarm setpoint for Alarm B. | R/W | Float | 0 |
rLowAlarmAlarmC | 40EC | Low Alarm C | Sets the Low Alarm setpoint for Alarm C. | R/W | Float | 0 |
rLowAlarmAlarmD | 40ED | Low Alarm D | Sets the Low Alarm setpoint for Alarm D. | R/W | Float | 0 |
rHighDeadbandAlarmA | 40EE | High Deadband A | Sets the High Alarm deadband for Alarm A. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint | R/W | Float | 0 |
rHighDeadbandAlarmB | 40EF | High Deadband B | Sets the High Alarm deadband for Alarm B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint | R/W | Float | 0 |
rHighDeadbandAlarmC | 40F0 | High Deadband C | Sets the High Alarm deadband for Alarm C. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint | R/W | Float | 0 |
rHighDeadbandAlarmD | 40F1 | High Deadband D | Sets the High Alarm deadband for Alarm D. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint | R/W | Float | 0 |
rLowDeadbandAlarmA | 40F2 | Low Deadband A | Sets the Low Alarm deadband for Alarm A. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint | R/W | Float | 0 |
rLowDeadbandAlarmB | 40F3 | Low Deadband B | Sets the Low Alarm deadband for Alarm B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint | R/W | Float | 0 |
rLowDeadbandAlarmC | 40F4 | Low Deadband C | Sets the Low Alarm deadband for Alarm C. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint | R/W | Float | 0 |
rLowDeadbandAlarmD | 40F5 | Low Deadband D | Sets the Low Alarm deadband for Alarm D. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint | R/W | Float | 0 |
rSetpointControlA | 40F6 | Setpoint Control A | Sets the Controller Setpoint for Relay A. | R/W | Float | 0 |
rSetpointControlB | 40F7 | Setpoint Control B | Sets the Controller Setpoint for Relay B. | R/W | Float | 0 |
rSetpointControlC | 40F8 | Setpoint Control C | Sets the Controller Setpoint for Relay C. | R/W | Float | 0 |
rSetpointControlD | 40F9 | Setpoint Control D | Sets the Controller Setpoint for Relay D. | R/W | Float | 0 |
rDeadbandControlA | 40FA | Sets the Controller Deadband/Proportional Band for Relay A. | R/W | Float | 0 | |
rDeadbandControlB | 40FB | Sets the Controller Deadband/Proportional Band for Relay B. | R/W | Float | 0 | |
rDeadbandControlC | 40FC | Sets the Controller Deadband/Proportional Band for Relay C. | R/W | Float | 0 | |
Name | ID | Label | Description | Type | Access | Class |