Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
eTestRelayValueB | 41A1 | Test Relay B Value | Sets the test state for Relay B. | R/W | Enum | 4 |
eTestRelayValueC | 41A2 | Test Relay C Value | Sets the test state for Relay C. | R/W | Enum | 4 |
eTestRelayValueD | 41A3 | Test Relay D Value | Sets the test state for Relay D. | R/W | Enum | 4 |
eDiscreteInput1Mode | 41A4 | Discrete Input 1 Mode | The Modes ON/High and OFF/Low are set by jumpers per the Instruction Manual. | R/W | Enum | 1024 |
eDiscreteInput2Mode | 41A5 | Discrete Input 2 Mode | The Modes ON/High and OFF/Low are set by jumpers per the Instruction Manual. | R/W | Enum | 1024 |
eDiscreteInput3Mode | 41A6 | Discrete Input 3 Mode | The Modes ON/High and OFF/Low are set by jumpers per the Instruction Manual. | R/W | Enum | 1024 |
eDiscreteInput1WarnEnable | 41A7 | DI 1 Warning Enable | When Enabled, a sc200 Warning will be generated when the Input is Active. | R/W | Enum | 1024 |
eDiscreteInput2WarnEnable | 41A8 | DI 2 Warning Enable | When Enabled, a sc200 Warning will be generated when the Input is Active. | R/W | Enum | 1024 |
eDiscreteInput3WarnEnable | 41A9 | DI 3 Warning Enable | When Enabled, a sc200 Warning will be generated when the Input is Active. | R/W | Enum | 1024 |
eDiscreteInput1HoldMode | 41AA | DI 1 Hold Mode | Selection list for the Discrete Input's Output Hold Mode. | R/W | Enum | 1024 |
eDiscreteInput2HoldMode | 41AB | DI 2 Hold Mode | Selection list for the Discrete Input's Output Hold Mode. | R/W | Enum | 1024 |
eDiscreteInput3HoldMode | 41AC | DI 3 Hold Mode | Selection list for the Discrete Input's Output Hold Mode. | R/W | Enum | 1024 |
iDiscreteInput1OnDelay | 41AD | DI 1 On Delay (0-999) | The delay in seconds before the Input goes Active. | R/W | Unsigned | 1024 |
iDiscreteInput2OnDelay | 41AE | DI 2 On Delay (0-999) | The delay in seconds before the Input goes Active. | R/W | Unsigned | 1024 |
iDiscreteInput3OnDelay | 41AF | DI 3 On Delay (0-999) | The delay in seconds before the Input goes Active. | R/W | Unsigned | 1024 |
iDiscreteInput1OffDelay | 41B0 | DI 1 Off Delay (0-999) | The delay in seconds before the Input goes Inactive. | R/W | Unsigned | 1024 |
iDiscreteInput2OffDelay | 41B1 | DI 2 Off Delay (0-999) | The delay in seconds before the Input goes Inactive. | R/W | Unsigned | 1024 |
iDiscreteInput3OffDelay | 41B2 | DI 3 Off Delay (0-999) | The delay in seconds before the Input goes Inactive. | R/W | Unsigned | 1024 |
iOneLowUnsignedLongIntegerIndexSC200 | 41B5 | R/W | Index | 4096 | ||
iNetworkModuleGoodMsg | 41B7 | Net Mod Good Msg | Number of good messages on the network module. | R | Unsigned | 1 |
iOneHighUnsignedLongIntegerIndexSC200 | 41B9 | R/W | Index | 4096 | ||
iNetworkModuleStatsBadMsg | 41BB | Net Mod Bad Msg | Number of bad messages on the Network Card Port. | R | Unsigned | 1 |
iServicePortStatsGoodMsg | 41BC | Svc Port Good Msg | Number of good messages on the Service Port. | R | Unsigned | 1 |
iServicePortStatsBadMsg | 41BD | Svc Port Bad Msg | Number of bad messages on the Service Port. | R | Unsigned | 1 |
iPort1StatsGoodMsg | 41BE | Port 1 Good Msg | Number of good messages on Port 1. | R | Unsigned | 1 |
iPort1StatsBadMsg | 41BF | Port 1 Bad Msg | Number of bad messages on Port 1. | R | Unsigned | 1 |
iPort2StatsGoodMsg | 41C0 | Port 2 Good Msg | Number of good messages on Port 2. | R | Unsigned | 1 |
iPort2StatsBadMsg | 41C1 | Port 2 Bad Msg | Number of bad messages on Port 2. | R | Unsigned | 1 |
iPort3StatsGoodMsg | 41C2 | Port 3 Good Msg | Number of good messages on Port 3. | R | Unsigned | 1 |
iPort3StatsBadMsg | 41C3 | Port 3 Bad Msg | Number of bad messages on Port 3. | R | Unsigned | 1 |
iPort4StatsGoodMsg | 41C4 | Port 4 Good Msg | Number of good messages on Port 4. | R | Unsigned | 1 |
iPort4StatsBadMsg | 41C5 | Port 4 Bad Msg | Number of bad messages on Port 4. | R | Unsigned | 1 |
sMathUnits | 41C7 | Math Units | Six character measurement unit for the sc200 Calculated Value. | R | ASCII | 2048 |
sMathParameters | 41C8 | Math Parameters | Six character measurement parameter for the sc200 Calculated Value. | R | ASCII | 2048 |
sDigitalSensor1Connected | 41DE | Sensor 1 Status | Sensor 1 Installation Status. | R | ASCII | 4096 |
sDigitalSensor2Connected | 41DF | Sensor 2 Status | Sensor 2 Installation Status | R | ASCII | 4096 |
rLoop1Current | 422A | Loop 1 Current | Schleifenstrom- Wert des Schleifenstroms so wie er vom Gerät erkannt wird. | R/W | Float | 66 |
rLoop2Current | 422B | Loop 2 Current | Schleifenstrom- Wert des Schleifenstroms so wie er vom Gerät erkannt wird. | R/W | Float | 66 |
rLoopACurrent | 4223 | Loop A Current | Schleifenstrom- Wert des Schleifenstroms so wie er vom Gerät erkannt wird. | R/W | Float | 66 |
eLoopAAlarmCode | 4224 | Alarm A Select | Alarmtyp des Analogausgangs- definiert das Verhalten des Analogausgangs, wenn das Feldgerät erkennt, dass der Analogausgang eventuell nicht mehr mit der Variable des Feldgeräts übereinstimmen kann. ANMERKUNG- Der digitale Wert kann unbestimmt sein. | R | Enum | 64 |
bLoopAFlags | 4225 | Loop A Flags | Analogkanal Flag- definiert die Charakteristik eines Analogkanals | R/W | BitEnum | 64 |
rLoopBCurrent | 4226 | Loop B Current | Schleifenstrom- Wert des Schleifenstroms so wie er vom Gerät erkannt wird. | R/W | Float | 66 |
eLoopBAlarmCode | 4227 | Alarm B Select | Alarmtyp des Analogausgangs- definiert das Verhalten des Analogausgangs, wenn das Feldgerät erkennt, dass der Analogausgang eventuell nicht mehr mit der Variable des Feldgeräts übereinstimmen kann. ANMERKUNG- Der digitale Wert kann unbestimmt sein. | R | Enum | 64 |
bLoopBFlags | 4228 | Loop B Flags | Analogkanal Flag- definiert die Charakteristik eines Analogkanals | R | BitEnum | 64 |
rLoopCCurrent | 4229 | Loop C Current | Schleifenstrom- Wert des Schleifenstroms so wie er vom Gerät erkannt wird. | R/W | Float | 66 |
eLoopCAlarmCode | 422F | Alarm C Select | Alarmtyp des Analogausgangs- definiert das Verhalten des Analogausgangs, wenn das Feldgerät erkennt, dass der Analogausgang eventuell nicht mehr mit der Variable des Feldgeräts übereinstimmen kann. ANMERKUNG- Der digitale Wert kann unbestimmt sein. | R | Enum | 64 |
bLoopCFlags | 4230 | Loop C Flags | Analogkanal Flag- definiert die Charakteristik eines Analogkanals | R | BitEnum | 64 |
cAnalogChannelIndex | 4231 | R/W | Index | 4096 | ||
iHARTLowFloatIndex | 4236 | R/W | Index | 4096 | ||
rMinimumSettingLoopA | 405E | LRV Loop A | Lower Range Value Loop A. | R/W | Float | 66 |