Name ID Label Description Type Access Class
rDeadbandControlD 40FD Sets the Controller Deadband/Proportional Band for Relay D. R/W Float 0
rSetpointEventA 40FE Setpoint Event A Sets the Event Controller Setpoint for Relay A. R/W Float 0
rSetpointEventB 40FF Setpoint Event B Sets the Event Controller Setpoint for Relay B. R/W Float 0
rSetpointEventC 4100 Setpoint Event C Sets the Event Controller Setpoint for Relay C. R/W Float 0
rSetpointEventD 4101 Setpoint Event D Sets the Event Controller Setpoint for Relay D. R/W Float 0
rDeadbandEventA 4102 Sets the Controller Deadband/Dead Zone for Relay B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal. R/W Float 0
rDeadbandEventB 4103 Sets the Controller Deadband/Dead Zone for Relay B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal R/W Float 0
rDeadbandEventC 4104 Sets the Controller Deadband/Dead Zone for Relay C. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal R/W Float 0
rDeadbandEventD 4105 Sets the Controller Deadband/Dead Zone for Relay D. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal R/W Float 0
rPIManualPWMA 4106 Manual Value A Sets the Manual control output level (%) Relay A. R/W Float 0