Loading…
Name ID Label Description Type Access Class
cX10 401C X10 R Unsigned 8
cY10 4027 Y10 R Unsigned 8
fLSignalCal 409E primary Low signal Low value for primary signal calibration. Must be between 3.5mA and 8.0mA . R/W Float 4096
fUSignalCal 409F primary high signal high value for primary signal calibration. Must be between 16.0mA and 21.0mA . R/W Float 4096
fLPresCal 4032 Low pressure Low value for pressure sensor calibration. Must be between 0 and 9psi / 0 and 0.6bar / 0 and 60kPa . R/W Float 4096
fUPresCal 4033 High pressure High value for pressure sensor calibration. Must be between 10 and 120psi / 0.7 and 8.0bar / 70 and 800kPa . R/W Float 4096
ucOption 4004 Option SVI II option detection R BitEnum 8
nAdress 4095 Adress W Unsigned 8
svi_cSwitch1Type 407D switch1 acting Define switch1 actions. R/W Enum 8
svi_cSwitch2Type 4080 switch2 acting Define switch2 actions. R/W Enum 8
Name ID Label Description Type Access Class