Name ID Label Description Type Access Class
press_A_family_status 458E Family status Device Family Status- Indicates additional status information on a device variable for a specific device family. R BitEnum 10
drive_signal 4033 Drive Signal Displayed in % of the maximum drive, the drive signal is the circuit board's output to the current-to-pressure (I/P) converter. R Float 2050
drive_signal_units 4034 Drive Signal Units R Enum 2048
drive_signal_limit_status 458F LS Limit Status- Indicates whether the device variable is limited (i.e., not responding to a process). R Enum 10
drive_signal_quality 4590 PDQ Process Data Quality- Indicates the overall process data quality of the device variable value. R Enum 10
drive_signal_class 4591 Class Classification- Indicates the unit code expansion table used for a particular dynamic or device variable. R Enum 2048
drive_signal_family_status 4592 Family status Device Family Status- Indicates additional status information on a device variable for a specific device family. R BitEnum 10
press_B 4035 Pressure B Displays the output pressure from Port B of the instrument. R Float 2050
press_B_limit_status 4593 LS Limit Status- Indicates whether the device variable is limited (i.e., not responding to a process). R Enum 10
press_B_quality 4594 PDQ Process Data Quality- Indicates the overall process data quality of the device variable value. R Enum 10
Name ID Label Description Type Access Class