Name ID Label Description Type Access Class
Function_Mode 401F Function mode Select between measurement using one or two sensor inputs. R/W Enum 2048
PV_Mode 4020 PV mode When measuring using two sensor inputs various calculation functions of the PV can be selected. If PV=SV1 is selected then sensor 2 can be individually set up. R/W Enum 2048
PV_Unit 4021 PV unit フィールド機器バリアブル単位 - 単位はデジタル値, 最小スパン, Upper レンジ値, Lower レンジ値, Lower センサー リミット, Upper センサー リミットの表示に使用されます。 R/W Enum 32
RJ_Max_Indicator 403F Max RJ value Shows the maximum RJ value. R Float 34
RJ_Min_Indicator 4040 Min RJ value Shows the minimum RJ value. R Float 34
Extended_Functions 4030 Drift alert mode Drift alarm: On exceeding the alarm set point an attention symbol appears in the display and the temperature transmitter goes into fault signal. Drift warning: On exceeding the alarm set point only the attention symbol appears in the display. R/W Enum 2048
Drift_Alert_Value 4031 Drift alert value Define the drift alarm set points for drift warning and -alarm. R/W Float 2048
Alarm_Ambient_Temp 400F Ambient alert Activate the ambient temperature alarm. If the parameter OFF is selected only a warning is issued on exceeding the specification. If ON is selected then an alarm is initiated when this value is exceeded. R/W Enum 2048
Filter_50_60_Hz 4033 Mains filter Select power filter, dependent on matching frequency of individual power networks. R/W Enum 2048
Corrosion_Detection 4032 Corros. detection Switch corrosion monitor on or off. R/W Enum 2048