Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
pulse_output_units | 1F4E | Tot. unit | Pulse Output Unit- Engineering units that the frequency and totalizer outputs will be displayed with. | R/W | Enum | 32 |
flow_alarm_select | 1F83 | AO Alrm typ | Analog Output Alarm Type- Defines how the Analog Output will respond when the Field Device detects that the Analog Output may not be tracking the associated Field Device Variable. NOTE- The Digital Value representation may not be determinate. | R | Enum | 64 |
flow_lower_range_value | 1F86 | PV LRV | Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. | R/W | Float | 32 |
flow_upper_range_value | 1F85 | PV URV | Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. | R/W | Float | 32 |
flow_damping_value | 1F87 | PV Damping | Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. | R/W | Float | 32 |
flow_sensor_serial_number | 1F82 | PV Snsr s/n | Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. | R/W | Unsigned | 2048 |
flow_upper_sensor_limit | 1F7F | PV USL | Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. | R/W | Float | 8 |
flow_lower_sensor_limit | 1F80 | PV LSL | Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. | R/W | Float | 8 |
pulse_scaling | 1F4F | Pulse scaling | Pulse Scaling Factor- Magnitude of process variable per pulse. | R/W | Float | 512 |