Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
customer_language | 402D | Language | Language - Language of local control. | R/W | Enum | 8 |
positive_zero_return | 402E | PZR | Positive Zero Return - all flow values are suppressed. | R/W | Enum | 32 |
customer_code | 402F | Customer Code | Customer Code - Safety code for local control and write protect for HART. Entering correct code disables write protection. | R/W | Integer | 8 |
code_entry | 4004 | Code Entry | Code Entry - Safety code for local control and write protect for HART. Entering correct code disables write protection, entering -1 enables write protection. | R/W | Integer | 32 |
self_check_switch | 4030 | Self check | Self checking - Controls self check function of the measuring amplifier. | R/W | Enum | 32 |
amplifier_sw_version | 4031 | SW AMP | Software Amplifier - Software version of amplifier module. | R | PackedAscii | 32 |
commodule_sw_version | 4032 | SW COM | Software Communication - Software version of communication module. | R | PackedAscii | 32 |
flow_noise_suppression | 4033 | Noise sup | Noise sup - Noise suppression is a software filter to reduce the sensitivity against transient flows and interference peaks. | R/W | Enum | 32 |
empty_pipe_control | 401B | EPD | EPD - Empty pipe detection control and calibration. | R/W | Enum | 34 |
empty_pipe_delay | 4035 | EPD Delay | EPD Delay - Delay off empty pipe detection. | R/W | Enum | 32 |