Loading…
Name ID Label Description Type Access Class
pressureLSL 400B Pres LSL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 4096
pressureMinimumSpan 400C Pres min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 4096
pressureDampingValue 400D Pres damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 4096
pressureSerialNumber 400E Pres snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R Unsigned 4096
pressureLowerTrimPoint 400F Pres Lower trim Lower Trim Point- Applied process input at which the last lower trim of the device variable was performed. R Float 4096
pressureUpperTrimPoint 4010 Pres Upper trim Upper Trim Point- Applied process input at which the last upper trim of the device variable was performed. R Float 4096
pressureTrimSupport 4011 Pres Trim support Trim Points Supported- Indicates the trim points that are supported by the device variable during a trim procedure. R Enum 4096
pressureMinimumLTP 4012 Pres Min lower trim Minimum Lower Trim Point- The lowest applied process value that can be used duirng a low trim procedure (ie, no value lower than this will be accepted by the instrument). R Float 4096
pressureMaximumLTP 4013 Pres Max lower trim Maximum Lower Trim Point- The highest applied process value that can be used duirng a low trim procedure (ie, no value higher than this will be accepted by the instrument). R Float 4096
pressureMinimumUTP 4014 Pres Min upper trim Minimum Upper Trim Point- The lowest applied process value that can be used duirng a high trim procedure (ie, no value lower than this will be accepted by the instrument). R Float 4096
Name ID Label Description Type Access Class