Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
pressureLSL | 400B | Pres LSL | Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. Pressure- Pressure of the process measured with respect to a reference pressure. | R | Float | 4096 |
pressureMinimumSpan | 400C | Pres min span | Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. Pressure- Pressure of the process measured with respect to a reference pressure. | R | Float | 4096 |
pressureDampingValue | 400D | Pres damp | Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. Pressure- Pressure of the process measured with respect to a reference pressure. | R | Float | 4096 |
pressureSerialNumber | 400E | Pres snsr s/n | Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. | R | Unsigned | 4096 |
pressureLowerTrimPoint | 400F | Pres Lower trim | Lower Trim Point- Applied process input at which the last lower trim of the device variable was performed. | R | Float | 4096 |
pressureUpperTrimPoint | 4010 | Pres Upper trim | Upper Trim Point- Applied process input at which the last upper trim of the device variable was performed. | R | Float | 4096 |
pressureTrimSupport | 4011 | Pres Trim support | Trim Points Supported- Indicates the trim points that are supported by the device variable during a trim procedure. | R | Enum | 4096 |
pressureMinimumLTP | 4012 | Pres Min lower trim | Minimum Lower Trim Point- The lowest applied process value that can be used duirng a low trim procedure (ie, no value lower than this will be accepted by the instrument). | R | Float | 4096 |
pressureMaximumLTP | 4013 | Pres Max lower trim | Maximum Lower Trim Point- The highest applied process value that can be used duirng a low trim procedure (ie, no value higher than this will be accepted by the instrument). | R | Float | 4096 |
pressureMinimumUTP | 4014 | Pres Min upper trim | Minimum Upper Trim Point- The lowest applied process value that can be used duirng a high trim procedure (ie, no value lower than this will be accepted by the instrument). | R | Float | 4096 |
Name | ID | Label | Description | Type | Access | Class |