Loading…
Name ID Label Description Type Access Class
output_transfer_function 4001 Xfer fnctn R Enum 0
dummy_unit 4002 Dummy unit R/W Enum 4096
dummy_float 4003 Dummy float R/W Float 4096
pressureValue 4004 Pres Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 4096
pressureFamily 4005 Pres Dev fmly Device Family- Indicates the family (i.e, pressure, temperature, flow, etc.) that the device variable belongs. R Enum 4096
pressureClassification 4006 Pres Class Classification- Indicates the unit code expansion table used for a particular dynamic or device variable. R Enum 4096
pressurePDQ 4007 Pres PDQ Process Data Quality- Indicates the overall process data quality of the device variable value. R Enum 4096
pressureLS 4008 Pres LS Limit Status- Indicates whether the device variable is limited (i.e., not responding to a process). R Enum 4096
pressureStatus 4009 Pres Fmly stat Device Family Status- Indicates additional status information on a device variable for a specific device family. R BitEnum 4096
pressureUSL 400A Pres USL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 4096
pressureLSL 400B Pres LSL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 4096
pressureMinimumSpan 400C Pres min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 4096
pressureDampingValue 400D Pres damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 4096
pressureSerialNumber 400E Pres snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R Unsigned 4096
pressureLowerTrimPoint 400F Pres Lower trim Lower Trim Point- Applied process input at which the last lower trim of the device variable was performed. R Float 4096
pressureUpperTrimPoint 4010 Pres Upper trim Upper Trim Point- Applied process input at which the last upper trim of the device variable was performed. R Float 4096
pressureTrimSupport 4011 Pres Trim support Trim Points Supported- Indicates the trim points that are supported by the device variable during a trim procedure. R Enum 4096
pressureMinimumLTP 4012 Pres Min lower trim Minimum Lower Trim Point- The lowest applied process value that can be used duirng a low trim procedure (ie, no value lower than this will be accepted by the instrument). R Float 4096
pressureMaximumLTP 4013 Pres Max lower trim Maximum Lower Trim Point- The highest applied process value that can be used duirng a low trim procedure (ie, no value higher than this will be accepted by the instrument). R Float 4096
pressureMinimumUTP 4014 Pres Min upper trim Minimum Upper Trim Point- The lowest applied process value that can be used duirng a high trim procedure (ie, no value lower than this will be accepted by the instrument). R Float 4096
pressureMaximumUTP 4015 Pres Max upper trim Maximum Upper Trim Point- The highest applied process value that can be used duirng a high trim procedure (ie, no value higher than this will be accepted by the instrument). R Float 4096
pressureMinimumTrimDiff 4016 Pres Min trim diff Minimum Trim Differential- The minimum acceptable difference between the upper and lower trim points. R Float 4096
AccelerationOfGravity 4035 Acc. of gravity R/W Float 32
AcceptPressureLowerRangeValue 4037 Get LRV Sets lower range value / Reference pressure is present at device R/W Enum 32
AcceptPressureUpperRangeValue 4038 Get URV Sets upper range value / Reference pressure is present at device R/W Enum 32
Name ID Label Description Type Access Class