Loading…
Name ID Label Description Type Access Class
Snsr0CalibSerialNumber 40C7 Conc snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R/W Unsigned 2048
Snsr0CalibSimulated 40C8 Simulate W Enum 2052
Snsr0CalibLowerTrimPoint 40C9 Conc Lower trim Lower Trim Point- Applied process input at which the last lower trim of the device variable was performed. R Float 0
Snsr0CalibUpperTrimPoint 40CA Conc Upper trim Upper Trim Point- Applied process input at which the last upper trim of the device variable was performed. R Float 0
Snsr0CalibTrimSupport 40CB Conc Trim support Trim Points Supported- Indicates the trim points that are supported by the device variable during a trim procedure. R Enum 0
Snsr0CalibMinimumLTP 40CC Conc Min lower trim Minimum Lower Trim Point- The lowest applied process value that can be used duirng a low trim procedure (ie, no value lower than this will be accepted by the instrument). R Float 0
Snsr0CalibMaximumLTP 40CD Conc Max lower trim Maximum Lower Trim Point- The highest applied process value that can be used duirng a low trim procedure (ie, no value higher than this will be accepted by the instrument). R Float 0
Snsr0CalibMinimumUTP 40CE Conc Min upper trim Minimum Upper Trim Point- The lowest applied process value that can be used duirng a high trim procedure (ie, no value lower than this will be accepted by the instrument). R Float 0
Snsr0CalibMaximumUTP 40CF Conc Max upper trim Maximum Upper Trim Point- The highest applied process value that can be used duirng a high trim procedure (ie, no value higher than this will be accepted by the instrument). R Float 0
Snsr0CalibMinimumTrimDiff 40D0 Conc Min trim diff Minimum Trim Differential- The minimum acceptable difference between the upper and lower trim points. R Float 0
Name ID Label Description Type Access Class