Loading…
Name ID Label Description Type Access Class
Snsr3TemperatureUpperTrimPoint 41AE Temp Upper trim Upper Trim Point- Applied process input at which the last upper trim of the device variable was performed. R Float 0
Snsr3TemperatureTrimSupport 41AF Temp Trim support Trim Points Supported- Indicates the trim points that are supported by the device variable during a trim procedure. R Enum 0
Snsr3TemperatureMinimumLTP 41B0 Temp Min lower trim Minimum Lower Trim Point- The lowest applied process value that can be used duirng a low trim procedure (ie, no value lower than this will be accepted by the instrument). R Float 0
Snsr3TemperatureMaximumLTP 41B1 Temp Max lower trim Maximum Lower Trim Point- The highest applied process value that can be used duirng a low trim procedure (ie, no value higher than this will be accepted by the instrument). R Float 0
Snsr3TemperatureMinimumUTP 41B2 Temp Min upper trim Minimum Upper Trim Point- The lowest applied process value that can be used duirng a high trim procedure (ie, no value lower than this will be accepted by the instrument). R Float 0
Snsr3TemperatureMaximumUTP 41B3 Temp Max upper trim Maximum Upper Trim Point- The highest applied process value that can be used duirng a high trim procedure (ie, no value higher than this will be accepted by the instrument). R Float 0
Snsr3TemperatureMinimumTrimDiff 41B4 Temp Min trim diff Minimum Trim Differential- The minimum acceptable difference between the upper and lower trim points. R Float 0
ModbusDisplayCode 404F Port Select Selects port number R/W Enum 4096
CalibDisplayCode 41C2 Sensor Select Select sensor for menu functions R/W Enum 4096
AlarmDisplayCode 41C3 Sensor Alarm Limit Selects sensor for alarm menu functions R/W Enum 4096
Name ID Label Description Type Access Class