Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
Snsr3CalibUSL | 418F | Conc USL | Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. | R | Float | 0 |
Snsr3CalibLSL | 4190 | Conc LSL | Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. | R | Float | 0 |
Snsr3CalibMinimumSpan | 4191 | Conc min span | Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. | R | Float | 0 |
Snsr3CalibDampingValue | 4192 | Conc damp | Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. | R | Float | 0 |
Snsr3CalibUpdatePeriod | 4193 | Conc Update Period | R | Time | 0 | |
Snsr3CalibSerialNumber | 4194 | Conc snsr s/n | Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. | R/W | Unsigned | 2048 |
Snsr3CalibSimulated | 4195 | Simulate | W | Enum | 2052 | |
Snsr3CalibLowerTrimPoint | 4196 | Conc Lower trim | Lower Trim Point- Applied process input at which the last lower trim of the device variable was performed. | R | Float | 0 |
Snsr3CalibUpperTrimPoint | 4197 | Conc Upper trim | Upper Trim Point- Applied process input at which the last upper trim of the device variable was performed. | R | Float | 0 |
Snsr3CalibTrimSupport | 4198 | Conc Trim support | Trim Points Supported- Indicates the trim points that are supported by the device variable during a trim procedure. | R | Enum | 0 |