Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
conc0DampingValue | 4014 | Conc damp | Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. | R | Float | 0 |
conc0UpdatePeriod | 4015 | Conc Update Period | R | Time | 0 | |
conc0SerialNumber | 4016 | Conc snsr s/n | Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. | R/W | Unsigned | 2048 |
conc0Simulated | 4017 | Simulate | W | Enum | 2052 | |
conc0LowerTrimPoint | 4018 | Conc Lower trim | Lower Trim Point- Applied process input at which the last lower trim of the device variable was performed. | R | Float | 0 |
conc0UpperTrimPoint | 4019 | Conc Upper trim | Upper Trim Point- Applied process input at which the last upper trim of the device variable was performed. | R | Float | 0 |
conc0TrimSupport | 401A | Conc Trim support | Trim Points Supported- Indicates the trim points that are supported by the device variable during a trim procedure. | R | Enum | 0 |
conc0MinimumLTP | 401B | Conc Min lower trim | Minimum Lower Trim Point- The lowest applied process value that can be used duirng a low trim procedure (ie, no value lower than this will be accepted by the instrument). | R | Float | 0 |
conc0MaximumLTP | 401C | Conc Max lower trim | Maximum Lower Trim Point- The highest applied process value that can be used duirng a low trim procedure (ie, no value higher than this will be accepted by the instrument). | R | Float | 0 |
conc0MinimumUTP | 401D | Conc Min upper trim | Minimum Upper Trim Point- The lowest applied process value that can be used duirng a high trim procedure (ie, no value lower than this will be accepted by the instrument). | R | Float | 0 |
Name | ID | Label | Description | Type | Access | Class |