Name ID Label Description Type Access Class
pv_damp_value 4016 Damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R/W Float 2
gas_units 4001 PV unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R/W Enum 32
gas_value 4018 Gas level Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention R Float 2
gas_upper_sensor_limit 4019 PV USL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R/W Float 8
gas_lower_sensor_limit 401A PV LSL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R/W Float 8
gas_minimum_span 401B PV Min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R/W Float 32
gas_range_units 4000 PV rnge unit Range Unit- Engineering unit to be displayed with the Upper Range Value and Lower Range Value R/W Enum 32
cell_temp_units 4021 PV unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R/W Enum 32
cell_temp 4022 Cell Temperature Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention R Float 2
cell_voltage_units 401E PV unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R/W Enum 32
Name ID Label Description Type Access Class