Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
slope | 403E | Slope | Programmable constant. The value may be determined by the user | R/W | Float | 2048 |
limit1 | 403F | Limit1 | Alarm levels may be defined in OXITEC 5000. The levels are determined by the user | R/W | Float | 2048 |
limit2 | 4040 | Limit2 | Alarm levels may be defined in OXITEC 5000. The levels are determined by the user | R/W | Float | 2048 |
limit1_max_min | 4041 | Limit1 func | R/W | Enum | 2048 | |
limit2_max_min | 4042 | Limit2 func | R/W | Enum | 2048 | |
co_calib_zero_offset | 4043 | CO zero | Zero calibration offset for CO | R/W | Float | 2048 |
co_calib_span | 4044 | CO span | Span calibration value for CO | R/W | Float | 2048 |
co_measuring_range | 4045 | CO Range | Measuring Range for CO | R/W | Float | 2048 |
co_current | 4046 | CO Current | Loop current for CO | R | Float | 2048 |
device_status | 0097 | Device status | Field Device Status- Data indicating that the Field Device has detected condition(s) in the Field Device relating to: its hardware / the validity of a variable / its operating status / an internal process | R | BitEnum | 131 |