Loading…
Name ID Label Description Type Access Class
calInterval 4008 Cal Intv'l The sensor is scheduled to be calibrated this often R/W Unsigned 8
vRangeRef 406A RangeRef R Float 0
vAbsResolution 4069 AbsRes displays the Absolute Resolution R Float 0
vA1Level 404D Alarm1 Level Indicates the reading where Alarm1 is tripped R/W Float 0
vA1Band 404E Alarm1 Hysteresis Indicates the Hysteresis associated with Alarm1. Once a Rising Alarm is tripped is will not reset until the reading is less then (Alarm1 Level - Hysteresis) R/W Float 0
vA1Rising 4051 Alarm1 Direction If Rising, Alarm1 is tripped when the reading exceeds the Alarm1 Level R/W Enum 0
vA1Ack 4050 Alarm1 acknowledgable Indicates whether Alarm1 can be reset by the operator acknowledging it. R/W Enum 0
vA1Latch 404F Alarm1 latching Indicates whether Alarm1 stays tripped until reset by the operator acknowledging it. R/W Enum 0
vA2Level 4052 Alarm2 Level Indicates the reading where Alarm2 is tripped R/W Float 0
vA2Band 4053 Alarm2 Hysteresis Indicates the Hysteresis associated with Alarm2.Once a Rising Alarm is tripped is will not reset until the reading is less then (Alarm2 Level - Hysteresis) R/W Float 0
Name ID Label Description Type Access Class