Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
SENSOR_LOCK_INFO | 40A9 | Sensor Lock Info | If the SENSOR_LOCK_INFO parameter is activated, successful replacement of the sensors is only possible with identical part numbers. If the part number of the old sensor and the part number of the new sensor are not identical, an error message is issued. If, however, the SENSOR_LOCK_INFO parameter is deactivated, sensors can be replaced with sensors of the same sensor type with different part numbers. | R/W | Enum | 0 |
SENSOR_LOCK_PART_NUMBER | 40AA | Sensor Lock Part Number | Contains the part number (Draeger order number) of the sensor that can be replaced while SENSOR_LOCK_INFO is active | R | ASCII | 8 |
SENSOR_PART_NUMBER | 40AB | Part Number | Part number of the connected sensor | R | ASCII | 8 |
SENSOR_SN | 40AC | Serial Number | Serial number of the connected sensor | R | ASCII | 8 |
SENSOR_ID | 40AD | Sensor ID | R/W | Enum | 0 | |
SENSOR_VITALITY | 40AF | Sensor Vitality | SENSOR VITALITY | R | Enum | 10 |
CALIBRATION_POINT_NUM | 40B2 | Calibration Index | Calibration Index | R/W | Unsigned | 8 |
PERFORM_CALIBRATION_STATUS | 40B5 | Perform Calibration Status | Perform Calibration Status | R | Enum | 10 |
device_status | 0097 | Status de dispositivo | Status do dispositivo de campo- O dado indica que o dispositivo de campo detectou condição (s) no dispositivo de campo em relação à: seu hardware / a validade da variável / seu status de operação / um processo interno | R | BitEnum | 131 |
units_code | 07E9 | R/W | Enum | 4096 | ||
Name | ID | Label | Description | Type | Access | Class |