Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
CAL_UNIT_NAME | 408C | Cal Unit Name | Contains the name of the currently set target gas unit in ASCII format. | R | ASCII | 8 |
CAL_UNIT_INDEX | 408E | Cal Unit Index | Contains the index of the currently set target gas unit. The parameter can be modified and the respective target gas unit set by writing in the parameter. | R/W | Enum | 0 |
ALARM_LEVEL | 4092 | Alarm LeveL | Alarm level | R | Enum | 10 |
HIGHEST_SUPPORTED_ALARM_LEVEL | 4093 | Highest Supported Alarm Level | Highest supported alarm level | R | Unsigned | 8 |
A1_ALARM_MODE | 4094 | Alarm Mode | Alarm Mode | R | BitEnum | 10 |
A2_ALARM_MODE | 4095 | Alarm Mode | Alarm Mode | R | BitEnum | 10 |
A1_ALARM_CONFIGURATION_VALUE | 4096 | Level | Contains the currently set gas concentration for the A1 alarm. | R/W | Float | 10 |
A1_ALARM_CONFIGURATION_HYSTERESE | 4097 | Hysterese | The alarm hysteresis defines a range around the A1 alarm within which A1_ALARM remains set until the current gas concentration falls below this threshold. | R/W | Float | 10 |
A2_ALARM_CONFIGURATION_VALUE | 4098 | Level | Contains the currently set gas concentration for the A2 alarm. | R/W | Float | 10 |
A2_ALARM_CONFIGURATION_HYSTERESE | 4099 | Hysterese | The alarm hysteresis defines a range around the A2 alarm within which A2_ALARM remains set until the current gas concentration falls below this threshold. | R/W | Float | 10 |
Name | ID | Label | Description | Type | Access | Class |