Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
ZeroFlowCalReqControl | 401F | ZeroFlowCalReqControl | Specifies the desired zero-flow calibration action. A zero-flow calibration process must be inactive in order to be started. The "Accept Result and Exit" selection is only valid when the status indicates "Completed Successfully." | R/W | Enum | 8 |
ZeroFlowCalStatus | 401C | Zero calib status | Current zero-flow calibration status indicator. | R | Enum | 2 |
ZeroFlowCalReqDuration | 401E | Zero calib duration | Specifies the zero-flow calibration duration. | R/W | Unsigned | 0 |
DryCalVel | 404D | Zero calib velocity | Factory-calibrated flow velocity (customer calibration not applied). This is the result of applying the factory calibration coefficients to the average weighted flow velocity (AvgWtdFlowVel). | R | Float | 2 |
ZeroFlowCalResult | 4021 | Proposed offset | Zero-flow calibration result. This value is only valid and can only be accepted when the ZeroFlowCalStatus is "Completed Successfully." When accepted, this value is used to update the FwdA0 and RevA0 data points. This value is not retained through a power cycle. | R | Float | 2 |
ZeroFlowCalProgress | 404E | Zero calib progress | Zero-flow calibration progress (percent complete). | R | Unsigned | 2 |
HARTDevVar0Value | 404F | Uncorr volume flow rate | HART Device Variable 0 (uncorrected volumetric flow rate) value. The units are specified by the 'HARTAO2VolFlowRateUnit' data point. | R | Float | 2 |
HARTDevVar0XdcrLmtUpper | 4050 | HARTDevVar0XdcrLmtUpper | Specifies the HART Device Variable 0 (uncorrected volumetric flow rate) upper transducer limit. The units are specified by the 'HARTDevVar0Units' data point. | R | Float | 8 |
HARTDevVar0XdcrLmtLower | 4051 | HARTDevVar0XdcrLmtLower | Specifies the HART Device Variable 0 (uncorrected volumetric flow rate) lower transducer limit . The units are specified by the 'HARTDevVar0Units' data point. | R | Float | 8 |
HARTDevVar0MinimumSpan | 4052 | Device variable 0 minimum span | R | Float | 0 | |
Name | ID | Label | Description | Type | Access | Class |