名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
write_totalizer_TotDayReset | 405C | Tot Day reset | Resetting the totaliser in main flow direction (daily counter) | 写 | 浮点 | 2048 |
write_totalizer_TotAllReset | 405D | Tot All reset | Resetting the totaliser in main flow direction (total counter) | 写 | 浮点 | 2048 |
totalizer_TotPlus | 400E | Tot+ | Totaliser in main flow direction (total counter) | 读 | 浮点 | 2050 |
totalizer_TotMinus | 400F | Tot- | Totaliser reverse the main flow direction | 读 | 浮点 | 2050 |
totalizer_TotDay | 4010 | Tot Day | Totaliser in main flow direction (daily counter) | 读 | 浮点 | 2050 |
analog_calibration_offset | 4032 | Analog Out. Offset | Calibration of the analog output (already done at the factory). Adjusting the offset at 4 mA. | 读/写 | 无符号 | 2048 |
analog_calibration_gain | 4033 | Analog Out. Gain | Calibration of the analog output (already done at the factory). Adjusting the range at 20 mA. | 读/写 | 无符号 | 2048 |
empty_pipe_calibration_value | 4028 | Empty-Pipe-Value | Measuring signal at fulfilled pipe. | 读 | 无符号 | 2048 |
empty_pipe_calibration_hysteresisUnit | 4085 | Empty-Pipe-H. Unit | Measuring signal unit in mV | 读/写 | 枚举 | 2048 |
empty_pipe_calibration_hysteresis | 4029 | Empty-Pipe-Hyst. | Adjusting the hysteresis value (recommended appr. 1000 mV) | 读/写 | 无符号 | 2048 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |