Loading…
名称 ID 标签 描述 类型 访问 分类
volume_unit 4006 Volume Unit Adjusting the volume unit for totalisation and preset 读/写 枚举 2048
delete_error_list 405E delete Resetting of error list 无符号 2048
power_on_counter 4014 Power on counter This value indicates the total switching ons of the meter 无符号 2048
pulse_type 400C Pulse Type Adjusting the pulse type. Closed (no pulse = low / pulse = high) as well as open (no pulse = high / pulse = low) 读/写 枚举 2048
end_flow_simultion 4058 End Flow Simulation Flow simulation of all outputs with the adjusted parameters 无符号 2048
start_flow_simultion 4052 Start Flow Simulation Start flow simulation 无符号 2048
simulate_flow_plus100_percent 4053 Simulate Flow +100% 100% of the full scale in main flow direction as well as relay 1 (forward) 无符号 2048
simulate_flow_plus50_percent 4054 Simulate Flow +50% 50% of full scale in main flow direction as well as relay 1 (forward) + relay 2 无符号 2048
simulate_flow_plus0_percent 4055 Simulate Flow 0% 0% of full scale in main flow direction as well as relay 1 (forward) + relay 3 无符号 2048
simulate_flow_minus50_percent 4056 Simulate Flow -50% 50% of the full scale reverse the main flow direction as well as relay 1 (backward) + relay 2 无符号 2048
名称 ID 标签 描述 类型 访问 分类