Loading…
Name ID Label Description Type Access Class
var_upper_range_value 400A Up.Range Upper Range Limit R/W Float 32
var_lower_range_value 400B Lo.Range Lower Range Limit R/W Float 32
var_sensor_serial_number 400C SensorNo Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R/W Unsigned 2048
var_cps_average_unit 400D Cps Pulse Rate Average Unit R Enum 8
var_cps_average_value 400E Cps av Shows the average of the pulse rate, depending on the time constant R Float 10
loc_cps_average_value 40F4 Cps av Shows the average of the pulse rate, depending on the time constant R Float 4096
var_cps_average_upper_limit 400F Up.Limit Set up the upper limit of the pulse rate. The measurement will stop when the live pulse rate will exceed this limit. R/W Float 8
var_cps_average_lower_limit 4010 Lo.Limit Set up the lower limit of the pulse rate. The measurement will stop when the live pulse rate will under-run this limit. R/W Float 8
var_cps_sum_unit 4011 None Unit of live pluse rate (sum of all probes) R Enum 8
var_cps_sum_value 4012 Cps Shows the the live pulse rate (sum of all probes) R Float 10
Name ID Label Description Type Access Class