名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
var_no_of_detectors | 401E | # of Probes | Select(ed) total number of Main and Auxiliary Units used in that measurement | 读/写 | 无符号 | 2048 |
var_detector_id | 401F | Probe No | Select probe ID in order to show up its raw data. No. 0 shows always the Main Unit. No. 1,2,.. shows the Auxiliary Units. | 读/写 | 无符号 | 2048 |
var_detector_code | 4020 | Code | Probe code for automatic HV-control | 读/写 | 无符号 | 2048 |
var_detector_address | 4021 | Device ID | Auxiliary address of Probe <=> Probe id | 读/写 | 无符号 | 2048 |
var_detector_count_rate_main | 4022 | Cps | Shows the live pulse rate of selected Probe | 读 | 浮点 | 2050 |
var_detector_count_rate_aux | 4023 | Cps AUX | Probe pluse rate in Aux. channel | 读 | 浮点 | 2050 |
var_detector_temp | 4024 | Temp. | Probe Temperature | 读 | 浮点 | 2050 |
var_temp_units | 4025 | Temp.Unit | Select the temperature unit for the probe temperature. | 读/写 | 枚举 | 2048 |
var_detector_hv_ctrl | 4026 | HV Setup | HV-control set up: 0 = automatic HV; 300V - 1300V = manual HV | 读/写 | 无符号 | 2048 |
var_detector_hv_value | 4027 | HV Reading | Probe high voltage value | 读 | 无符号 | 2050 |