Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_dd_fSensorValue | 4000 | Live Rate | Display of Active unaveraged ReadIn rate.(Cps live) | R | Float | 2050 |
var_dd_fPV1 | 4003 | PV | Digital representation of measured process value(PV->Primary Variable). | R/W | Float | 10 |
var_dd_eDeviceStatus | 4007 | Device Status | Represents the Status of the Field Device in accordance to NE107 Normal Valid output signal. Maintenance required(M) Still valid output signal. Out of specification(S) Signal out of specified range. Function check(C) Temporary non-valid output signal. Failure(F) Non-valid output signal. | R | Enum | 10 |
var_dd_eCurrentLoopStatus | 4008 | Current AO Status | Actual Current Loop Status | R | Enum | 10 |
var_dd_eDataQuality | 400B | Data Quality | Process Data Quality- Indicates the overall process data quality of the device variable value. | R | Enum | 10 |
var_dd_eLimitStatus | 400C | Limit Status | Limit Status- Indicates whether the device variable is limited (i.e., not responding to a process). | R | Enum | 10 |
var_dd_beDeviceFamilyStatus | 400D | Family Status | Device Family Status- Indicates additional status information on a device variable for a specific device family. | R | BitEnum | 10 |
var_dd_fDensitySuspension | 400E | Density Suspension | R | Float | 10 | |
var_dd_fCpsZero | 400F | Io Rate | R | Float | 10 | |
var_dd_eMeasurementStatus | 4010 | Measurement Status | Display of actual measurement status. | R | Enum | 10 |
var_dd_fMeasPointRate1 | 4011 | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_dd_fMeasPointRate2 | 4012 | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_dd_fMeasPointRate3 | 4013 | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_dd_fMeasPointRate4 | 4014 | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_dd_fMeasPointRate5 | 4015 | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_dd_fMeasPointRate6 | 4016 | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_dd_fMeasPointRate7 | 4017 | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_dd_fMeasPointRate8 | 4018 | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_dd_fMeasPointRate9 | 4019 | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_dd_fMeasPointRate10 | 401A | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_dd_fMeasPointRate11 | 401B | Meas Rate | Rate for adjustment of meas point. | R | Float | 10 |
var_ds_uiOperatingHours | 401C | Operating Hours | Operating Hours- The counter value represents the accumulated release time of operation in hours | R/W | Unsigned | 34 |
var_dd_dtActiveDate | 401E | Date | Actual system date (mm/dd/yyyy). | R/W | HartDate | 10 |
var_dd_ascActiveTime | 401F | Time | Actual system time (hh:mm:ss). | R/W | ASCII | 10 |
var_dd_eOperatingMode | 4020 | Operating Mode | Displays actual System Operating Mode. | R | Enum | 2051 |
var_dd_uiInstructDescriptor | 4021 | Instruction Code | R | Enum | 3 | |
var_dd_uiErrorDescriptor | 4022 | Event Code | Event Code which is actually signaled by the Field device. | R | Enum | 3 |
var_dd_ucErrorToAcknowledge | 4023 | R | Enum | 8 | ||
var_dd_fDeviceTemperature | 4024 | Detector Temp. | Actual detector temperature value. | R | Float | 2050 |
var_dd_eTempMode | 4025 | Detector Temp. Mode | Select detector temperature mode. Select 'Fix Value' for test. | R/W | Enum | 2050 |
var_dd_fTempMin | 4026 | Detector Temp. Min | Minimum device temperature since last reset. | R | Float | 2050 |
var_dd_fTempMax | 4027 | Detector Temp. Max | Maximum device temperature since last reset. | R | Float | 2050 |
var_dd_fHVSetpoint | 4028 | HV Setpoint | Display of actual HV Setpoint at multiplier. | R | Float | 2050 |
var_dd_fHVMonitor | 4029 | HV Feedback | Display of actual HV feedback at multiplier. | R | Float | 2050 |
var_dd_fHVAverage | 402A | HV Average | Display HV Average as detector raw value. HV Average has a damping value of 10 days. The value is used for error messages 'HV-Error' and 'Change Multiplier'. | R | Float | 2050 |
var_dd_fProbeCps | 402B | Live Rate | Display of detector live count rate. | R | Float | 2050 |
var_dd_fMeasurementCps | 402C | Meas Channel | R | Float | 2050 | |
var_dd_fControlCps | 402D | Control Channel | R | Float | 2050 | |
var_dd_fAuxiliaryCps | 402E | Auxiliary Channel | R | Float | 2050 | |
var_dd_eHVMode | 402F | HV Mode | Display actual HV mode | R/W | Enum | 2050 |
var_dd_eDOValue | 4030 | Relay State | Relay State | R | Enum | 1026 |
var_dd_eLoopState | 4031 | Current Mode | A fixed current value is simulated. | R | Enum | 2050 |
var_dd_fCpsAverage | 4032 | Average Rate | The ReadIn rate is newly averaged in every measurement cycle. Standardized to one second. | R | Float | 10 |
var_dd_dtCalCompDate | 4033 | Compensation Date | Date of last compensation (mm/dd/yyyy). | R | HartDate | 10 |
var_dd_fReadInRemainedTime | 4034 | Remaining Time | Time remaining while ReadIn. | R | Float | 10 |
var_dd_fReadInCalBackground | 4035 | Background | Enter backgound radiation count rate. | R | Float | 2 |
var_dd_fReadInCalPointRate | 4036 | Cal Rate | Rate for adjustment of Cal Point. | R | Float | 2 |
var_dd_eDynamicDeviceSetup | 4037 | R | BitEnum | 10 | ||
var_dd_eDownloadStatus | 4038 | R | Enum | 2048 | ||
var_dd_eUploadStatus | 4039 | R | Enum | 2048 | ||
Name | ID | Label | Description | Type | Access | Class |