Loading…
Name ID Label Description Type Access Class
var_dd_fSensorValue 4000 Live Rate Display of Active unaveraged ReadIn rate.(Cps live) R Float 2050
var_dd_fPV1 4003 PV Digital representation of measured process value(PV->Primary Variable). R/W Float 10
var_dd_eDeviceStatus 4007 Device Status Represents the Status of the Field Device in accordance to NE107 Normal Valid output signal. Maintenance required(M) Still valid output signal. Out of specification(S) Signal out of specified range. Function check(C) Temporary non-valid output signal. Failure(F) Non-valid output signal. R Enum 10
var_dd_eCurrentLoopStatus 4008 Current AO Status Actual Current Loop Status R Enum 10
var_dd_eDataQuality 400B Data Quality Process Data Quality- Indicates the overall process data quality of the device variable value. R Enum 10
var_dd_eLimitStatus 400C Limit Status Limit Status- Indicates whether the device variable is limited (i.e., not responding to a process). R Enum 10
var_dd_beDeviceFamilyStatus 400D Family Status Device Family Status- Indicates additional status information on a device variable for a specific device family. R BitEnum 10
var_dd_fDensitySuspension 400E Density Suspension R Float 10
var_dd_fCpsZero 400F Io Rate R Float 10
var_dd_eMeasurementStatus 4010 Measurement Status Display of actual measurement status. R Enum 10
var_dd_fMeasPointRate1 4011 Meas Rate Rate for adjustment of meas point. R Float 10
var_dd_fMeasPointRate2 4012 Meas Rate Rate for adjustment of meas point. R Float 10
var_dd_fMeasPointRate3 4013 Meas Rate Rate for adjustment of meas point. R Float 10
var_dd_fMeasPointRate4 4014 Meas Rate Rate for adjustment of meas point. R Float 10
var_dd_fMeasPointRate5 4015 Meas Rate Rate for adjustment of meas point. R Float 10
var_dd_fMeasPointRate6 4016 Meas Rate Rate for adjustment of meas point. R Float 10
var_dd_fMeasPointRate7 4017 Meas Rate Rate for adjustment of meas point. R Float 10
var_dd_fMeasPointRate8 4018 Meas Rate Rate for adjustment of meas point. R Float 10
var_dd_fMeasPointRate9 4019 Meas Rate Rate for adjustment of meas point. R Float 10
var_dd_fMeasPointRate10 401A Meas Rate Rate for adjustment of meas point. R Float 10
var_dd_fMeasPointRate11 401B Meas Rate Rate for adjustment of meas point. R Float 10
var_ds_uiOperatingHours 401C Operating Hours Operating Hours- The counter value represents the accumulated release time of operation in hours R/W Unsigned 34
var_dd_dtActiveDate 401E Date Actual system date (mm/dd/yyyy). R/W HartDate 10
var_dd_ascActiveTime 401F Time Actual system time (hh:mm:ss). R/W ASCII 10
var_dd_eOperatingMode 4020 Operating Mode Displays actual System Operating Mode. R Enum 2051
var_dd_uiInstructDescriptor 4021 Instruction Code R Enum 3
var_dd_uiErrorDescriptor 4022 Event Code Event Code which is actually signaled by the Field device. R Enum 3
var_dd_ucErrorToAcknowledge 4023 R Enum 8
var_dd_fDeviceTemperature 4024 Detector Temp. Actual detector temperature value. R Float 2050
var_dd_eTempMode 4025 Detector Temp. Mode Select detector temperature mode. Select 'Fix Value' for test. R/W Enum 2050
var_dd_fTempMin 4026 Detector Temp. Min Minimum device temperature since last reset. R Float 2050
var_dd_fTempMax 4027 Detector Temp. Max Maximum device temperature since last reset. R Float 2050
var_dd_fHVSetpoint 4028 HV Setpoint Display of actual HV Setpoint at multiplier. R Float 2050
var_dd_fHVMonitor 4029 HV Feedback Display of actual HV feedback at multiplier. R Float 2050
var_dd_fHVAverage 402A HV Average Display HV Average as detector raw value. HV Average has a damping value of 10 days. The value is used for error messages 'HV-Error' and 'Change Multiplier'. R Float 2050
var_dd_fProbeCps 402B Live Rate Display of detector live count rate. R Float 2050
var_dd_fMeasurementCps 402C Meas Channel R Float 2050
var_dd_fControlCps 402D Control Channel R Float 2050
var_dd_fAuxiliaryCps 402E Auxiliary Channel R Float 2050
var_dd_eHVMode 402F HV Mode Display actual HV mode R/W Enum 2050
var_dd_eDOValue 4030 Relay State Relay State R Enum 1026
var_dd_eLoopState 4031 Current Mode A fixed current value is simulated. R Enum 2050
var_dd_fCpsAverage 4032 Average Rate The ReadIn rate is newly averaged in every measurement cycle. Standardized to one second. R Float 10
var_dd_dtCalCompDate 4033 Compensation Date Date of last compensation (mm/dd/yyyy). R HartDate 10
var_dd_fReadInRemainedTime 4034 Remaining Time Time remaining while ReadIn. R Float 10
var_dd_fReadInCalBackground 4035 Background Enter backgound radiation count rate. R Float 2
var_dd_fReadInCalPointRate 4036 Cal Rate Rate for adjustment of Cal Point. R Float 2
var_dd_eDynamicDeviceSetup 4037 R BitEnum 10
var_dd_eDownloadStatus 4038 R Enum 2048
var_dd_eUploadStatus 4039 R Enum 2048
Name ID Label Description Type Access Class