Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_dd_uiInstructDescriptor | 4021 | Instruction Code | R | Enum | 3 | |
var_dd_uiErrorDescriptor | 4022 | Event Code | Event Code which is actually signaled by the Field device. | R | Enum | 3 |
var_dd_ucErrorToAcknowledge | 4023 | R | Enum | 8 | ||
var_dd_fDeviceTemperature | 4024 | Detector Temp. | Actual detector temperature value. | R | Float | 2050 |
var_dd_eTempMode | 4025 | Detector Temp. Mode | Select detector temperature mode. Select 'Fix Value' for test. | R/W | Enum | 2050 |
var_dd_fTempMin | 4026 | Detector Temp. Min | Minimum device temperature since last reset. | R | Float | 2050 |
var_dd_fTempMax | 4027 | Detector Temp. Max | Maximum device temperature since last reset. | R | Float | 2050 |
var_dd_fHVSetpoint | 4028 | HV Setpoint | Display of actual HV Setpoint at multiplier. | R | Float | 2050 |
var_dd_fHVMonitor | 4029 | HV Feedback | Display of actual HV feedback at multiplier. | R | Float | 2050 |
var_dd_fHVAverage | 402A | HV Average | Display HV Average as detector raw value. HV Average has a damping value of 10 days. The value is used for error messages 'HV-Error' and 'Change Multiplier'. | R | Float | 2050 |
var_dd_fProbeCps | 402B | Live Rate | Display of detector live count rate. | R | Float | 2050 |
var_dd_fMeasurementCps | 402C | Meas Channel | R | Float | 2050 | |
var_dd_fControlCps | 402D | Control Channel | R | Float | 2050 | |
var_dd_fAuxiliaryCps | 402E | Auxiliary Channel | R | Float | 2050 | |
var_dd_eHVMode | 402F | HV Mode | Display actual HV mode | R/W | Enum | 2050 |
var_dd_eDOValue | 4030 | Relay State | Relay State | R | Enum | 1026 |
var_dd_eLoopState | 4031 | Current Mode | A fixed current value is simulated. | R | Enum | 2050 |
var_dd_fCpsAverage | 4032 | Average Rate | The ReadIn rate is newly averaged in every measurement cycle. Standardized to one second. | R | Float | 10 |
var_dd_dtCalCompDate | 4033 | Compensation Date | Date of last compensation (mm/dd/yyyy). | R | HartDate | 10 |
var_dd_fReadInRemainedTime | 4034 | Remaining Time | Time remaining while ReadIn. | R | Float | 10 |
var_dd_fReadInCalBackground | 4035 | Background | Enter backgound radiation count rate. | R | Float | 2 |
var_dd_fReadInCalPointRate | 4036 | Cal Rate | Rate for adjustment of Cal Point. | R | Float | 2 |
var_dd_eDynamicDeviceSetup | 4037 | R | BitEnum | 10 | ||
var_dd_eDownloadStatus | 4038 | R | Enum | 2048 | ||
var_dd_eUploadStatus | 4039 | R | Enum | 2048 | ||
Name | ID | Label | Description | Type | Access | Class |