Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
Pulse_width_value | 401B | Puls width | Pulse Width | R/W | Float | 512 |
Low_flow_cutoff | 402B | LFC value | Lo flo cutoff-Adjustable from 1 to 10% of the setting range. Below set value, output is driven to the zero flow rate signal level. Display is latched to zero. | R/W | Enum | 64 |
Flow_tube_size | 4010 | Tube size | Flow tube size- Nominal inside diameter of the pipe adjacent to the sensor. | R/W | Enum | 32 |
preset_value | 4031 | Totlizer cntrl | Totalizer Control Code- Controls the operation of all available totalizers. | R/W | Unsigned | 16 |
total_count | 4000 | Totlizer display | Totalizer display-Display totalized value. | R/W | Integer | 34 |
total_count_dmy | 4032 | Totlizer display | Totalizer display-Display totalized value. | R/W | Integer | 34 |
flow_gravity | 4005 | Gravity | Specific gravity | R/W | Float | 64 |
display_select | 400D | Disp select | Set main display. %: Percent of range. Rate: Flow rate. Total: Totalized value. | R/W | Enum | 64 |
ex_value | 4012 | EX value | EX value: Unique detector factor.Ex factor is stamped on the name plate. | R/W | Float | 32 |
flow_coefficient | 4008 | Coefficient | Set coefficient. Outputs are calculated by multiplying set coefficient. Default value is 1.0000. | R/W | Float | 32 |
Name | ID | Label | Description | Type | Access | Class |