Loading…
Name ID Label Description Type Access Class
ds_gascvd_sw_version 402E Operation S/W ver. Operation S/W ver.- GasCVD Operation software version number. R Unsigned 0
ds_oiml_sw_version 402F OIML S/W ver. OIML S/W ver.- GasCVD OIML R140 compliant software version number. R Unsigned 0
ds_t1t2 4030 Calculation parameter Calculation parameter- ISO t1/t2, t1(combustion temperature)/t2(metering temperature). GPA BP (dry/sat), base pressure and dry/sat condition. R Enum 0
ds_pv_output_code 412C PV output units PV output units- GasCVD PV output units R Enum 0
ds_oven_temp_sp 4031 Micro oven temp. SP Micro oven temp. SP- Sensor temperature Set Point. R Float 0
ds_rh60 4032 Rh Rh- Sensor resistance Rh. R Float 0
ds_rud60 4033 Rud Rud- Sensor resistance Rud. R Float 0
ds_ao_failsafe_dir 4034 Fail safe direction Fail safe direction- When a critical failure occurs, GasCVD sets the analog output to the fail safe (burnout) direction. Either Downscale (2.4 mA (-10%) ) or Upscale (21.6 mA (110%) ) can be selected. R/W Enum 0
ds_ao_failsafe_sim 4035 Set analog output Fail safe simulation R/W Enum 0
ds_ao_failsafe_sim_clear 4036 Clear analog output Fail safe simulation R/W Enum 0
Name ID Label Description Type Access Class