名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
ds_ao_high_limit | 4038 | Output high limit | Output high limit- Analog output current signal high limit. | 读 | 浮点 | 2048 |
ds_do_status | 4039 | STATUS output | STATUS output- Shows the STATUS output, Open or Close. | 读 | 枚举 | 2050 |
ds_do_status_type | 403A | STATUS type | STATUS type- Configures the STATUS output to OPEN or CLOSE at alarm condition. Normally Open (NO): When no alarm is detected, STATUS output is OPEN. When alarm is detected, STATUS output is CLOSE. Normally Close (NC): When no alarm is detected, STATUS output is CLOSE. When alarm is detected, STATUS output is OPEN. | 读/写 | 枚举 | 2048 |
ds_do_status_select | 403B | STATUS selection | STATUS selection- Select the STATUS output item from below. - None - Process alarm - Sensor temp. alarm - Calibration in process or Calibration failure - Process + Sensor temp. - Sensor temp. + Calibration - Calibration + Process - All | 读/写 | 枚举 | 2048 |
ds_do_status_sim | 403C | Set STATUS simulation | 读/写 | 枚举 | 2048 | |
ds_do_status_sim_clear | 403D | Clear STATUS simulation | 读/写 | 枚举 | 2048 | |
ds_do_status_sim_cont | 403E | Continue STATUS simulation | 读/写 | 枚举 | 2048 | |
ds_do_calib | 403F | CALIB output | CALIB output- Shows the CALIB output, Open or Close. | 读 | 枚举 | 2050 |
ds_do_calib_type | 4040 | CALIB type | CALIB type- The CALIB output is fixed at Normally Open (NO). Normally Open (NO): In normal operation, process gas will flow from Open port to GasCVD. And at the beginning of calibration, calibration gas will flow from Close port to GasCVD. | 读 | 枚举 | 2048 |
ds_do_calib_sim | 4041 | Set CALIB simulation | 读/写 | 枚举 | 2048 | |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |