Name ID Label Description Type Access Class
temperatureLSL 4081 Temp LSL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 0
temperatureMinimumSpan 4082 Temp min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 0
temperatureDampingValue 4083 Temp damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R Float 0
temperatureUpdatePeriod 4095 Temp Update Period R Time 0
temperatureSerialNumber 4085 Temp snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R Unsigned 2048
temperatureSimulated 4086 Simulate W Enum 2052
point_position 4016 Trim Mode Select trim mode R/W Enum 8
current_value 4018 Current trim value Current PV value R/W Float 2048
apply_value 4019 Apply trim value Change current value to apply value for trim R/W Float 2048
Dev_Xmtr_Var_Code 401B Xmtr Variable Device transmitter variable code W Enum 8