Name ID Label Description Type Access Class
current_lower_trim_point 4000 Lower current trim point Lower current trim point R Float 8
current_upper_trim_point 4001 Upper current trim point Upper current trim point R Float 8
sensor_lower_trim_point 4002 Lower sensor trim point Applied process pressure at which the last low sensor trim was performed. R/W Float 8
sensor_upper_trim_point 4003 Upper sensor trim point Applied process pressure at which the last upper sensor trim was performed. R/W Float 8
current_units 4004 Current units Current units R Enum 8
kind_of_trim 4005 Kind of trim User calibrations (sensor calibrations) are replaced by the original factory calibrations. W Enum 32
product_code 4006 Product code Product code as generated from the product catalogue. It is related to the pressure trasmitter and its optional. A modification of the product code will not affect device functionally. R ASCII 2048
separator_code 4007 Separator code Product code as generated from the product catalogue. It is related to the Remote Seal connected to the positive side of the transmitter. If no remote seal connected the code is blank. A modification of the product code will not affect device functionally. R ASCII 2048
manifold_code 4008 Manifold code Product code as generated from the product catalogue. It is related to the Remote Seal connected to the negative side of the transmitter.If no remote seal connected the code is blank. A modification of the product code will not affect device functionally R ASCII 2048
long_tag 4009 Long tag Tag 24 ASCII character R/W ASCII 2048
Name ID Label Description Type Access Class