Loading…
Name ID Label Description Type Access Class
uepSV 40A8 AO hi end pt Analog Output Upper Endpoint Value- Defines the upper endpoint, of the respective Analog Output, for normal operating conditions R Float 64
lepSV 40A9 AO lo end pt Analog Output Lower Endpoint Value- Defines the lower endpoint, of the respective Analog Output, for normal operating conditions R Float 64
ulvSV 40AA Upper Limit Value The maximum value of the analog signal. R Float 64
llvSV 40AB Lower Limit Value The minimum value of the analog signal. R Float 64
adpSV 40AC Added damp Analog Output Added Damping Value- Damping time constant applied the Analog Output. The units of this variable are always in seconds. R Float 64
acfSV 40AD Channel flags Analog Channel Flags- Defines the characteristics of the Analog Channel R BitEnum 64
pcrTV 40B0 TV % range Percent of range of the HART Secondary Variable (SV); tracks the associated measured value with respect to the Lower Range Value and Upper Range Value. R Float 34
tfcTV 40B1 Xfer fnctn R Enum 32
urvTV 40B2 URV Valore di campo superiore - Definisce il punto finale operativo da cui deriva il valore analogico e il punto 100% del campo percentuale. Inoltre, il valore di campo superiore definisce un punto finale operativo da cui derivano gli allarmi associati al valore analogico e gli allarmi associati alla rappresentazione del valore digitale. R/W Float 32
lrvTV 40B3 LRV Valore di campo inferiore - Definisce il punto finale operativo da cui si ricava il valore analogico e il punto 0% del campo percentuale. Inoltre, il valore di campo inferiore definisce un punto finale operativo da cui derivano gli allarmi associati al valore analogico e gli allarmi associati alla rappresentazione del valore digitale. R/W Float 32