Loading…
Name ID Label Description Type Access Class
difference_s2_s1_upper_sensor_limit 406A UpperLimit Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 0
difference_s2_s1_lower_sensor_limit 406B Lower Limit Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 0
difference_s2_s1_minimum_span 406C Min Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 0
snr2snr1qty 40DF Snr2-Snr1 Status Process Data Quality- Indicates the overall process data quality of the device variable value. R Enum 2
average_digital_units 406E Unit Unit for Sensor Averaging measurement of both Sensors and for all other relevant device variables R/W Enum 0
average_digital_value 406F Average Sensor 1 and Sensor 2 Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention R Float 2
damping_value_average 40E0 Damping Damping time constant applied to both the Digital Value representation and Analog Value of a respective Analog Output. The units of this variable are always in seconds R/W Float 0
average_upper_sensor_limit 4070 UpperLimit Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 0
average_lower_sensor_limit 4071 Lower Limit Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 0
average_minimum_span 4072 Min Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 0
Name ID Label Description Type Access Class