Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
primary_measurement_upper_sensor_limit | 401B | USL | Upper sensor limit value - Value that defines the absolute upper endpoint (largest magnitude) for the UPPER RANGE VALUE. | R | Float | 8 |
primary_measurement_lower_sensor_limit | 401C | LSL | Lower sensor limit value - Value that defines the absolute lower endpoint (smallest magnitude) for the LOWER RANGE VALUE. | R | Float | 8 |
primary_measurement_minimum_span | 401D | Intervalo mínimo | Minimum span value - Value that defines the smallest absolute value for the resulting figure derived from the following equation (UPPER RANGE VALUE - LOWER RANGE VALUE). | R | Float | 8 |
ma_analog_output_value | 4042 | AO | Saída analógica- saída física mensurável que altera em relação ao tempo, sem a intervenção do usuário | R/W | Float | 66 |
ma_analog_output_alarm_select | 4043 | Alarm type | Tipo de alarme de saída analógica- Define como a saída analógica responderá quando o dispositivo de campo detectar a saída analógica poderá não seguir a variáveldo dispositivo de campo. NOTA- A representação do valor digital poderá não ser determinada. | R | Enum | 64 |
electronics_temperature_digital_units | 4025 | Unidade de tempo CJC | Temp. unit of the reference junction temperature - The temperature on the terminal block of the field unit is always stated in °C during measurements with internal or external reference junction. | R | Enum | 34 |
electronics_temperature_digital_value | 4008 | Temperatura CJC | Reference junction temperature value - Temperature associated with the internal/external reference junction. At an internal reference junction the temperature of the connection terminals of the transmitter will be measured during measurements. | R/W | Float | 34 |
sensor1_digital_units | 401F | Unidades | Unit - physical unit in which the electrical value of the sensor is expressed. | R | Enum | 34 |
sensor1_measurement_digital_value | 4020 | Sensor 1 | Sensor value - electrical value of the sensor at the input. | R | Float | 34 |
sensor2_digital_units | 4022 | Unidades | Unit - physical unit in which the electrical value of the sensor is expressed. | R | Enum | 34 |
Name | ID | Label | Description | Type | Access | Class |