Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_TB82_cell_constant | 4010 | Cell Constant | Sensor cell constant | R/W | Float | 8 |
var_TB82_sensor_diagnostics | 4020 | Snsr diagnostics | Sensor Diagnostics selection determines whether certain sensor diagnostics are enabled or disabled | R/W | Enum | 8 |
var_TB82_usrdef_secondary | 4011 | Concen curve name | This is a six character text name for the user-defined concentration curve. | R/W | ASCII | 2048 |
var_TB82_usrdef_unit | 4012 | User def conc unit | Engineering units for user defined concentration curve | R/W | Enum | 8 |
var_TB82_concentration_range | 4013 | Concen range | The concentration range for the user defined concentration curve | R/W | Enum | 8 |
var_TB82_temp_sensor_type | 4021 | Temp snsr type | Temperature sensor type | R/W | Enum | 8 |
var_TB82_temp_comp_type | 4022 | Temp comp type | Determines the type of Temperature Compensation algorithm applied to the Primary Value | R/W | Enum | 8 |
var_TB82_manual_temp_setpoint_unit | 4042 | Manual temp unit | R/W | Enum | 8 | |
var_TB82_manual_temp_setpoint | 4027 | Manual Temp. | Manual Temperature value is used in the temp comp algorithm when the temp comp type is MANUAL. | R/W | Float | 8 |
var_TB82_temp_auto_sol | 4023 | Auto TC option | The Automatic Temperature Compensation option applies an algorithm that adjusts for the temperature effect on conductivity, which is dependent on the selected solute. The available options are: Std KCl, TC Coeff, Pure H2O, User Defined. | R/W | Enum | 8 |
Name | ID | Label | Description | Type | Access | Class |