Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_TB82_temp_sensor_type | 4021 | Temp snsr type | Temperature sensor type | R/W | Enum | 8 |
var_TB82_temp_comp_type | 4022 | Temp comp type | Determines the type of Temperature Compensation algorithm applied to the Primary Value | R/W | Enum | 8 |
var_TB82_manual_temp_setpoint_unit | 4042 | Manual temp unit | R/W | Enum | 8 | |
var_TB82_manual_temp_setpoint | 4027 | Manual Temp. | Manual Temperature value is used in the temp comp algorithm when the temp comp type is MANUAL. | R/W | Float | 8 |
var_TB82_temp_auto_sol | 4023 | Auto TC option | The Automatic Temperature Compensation option applies an algorithm that adjusts for the temperature effect on conductivity, which is dependent on the selected solute. The available options are: Std KCl, TC Coeff, Pure H2O, User Defined. | R/W | Enum | 8 |
var_TB82_temp_pure_h2o | 4024 | Auto TC pure H2O options | This is valid when temp-comp type is Automatic and Auto TC option is Pure H2O. This parameter provides Pure H2O option selection for temp-comp. | R/W | Enum | 8 |
var_TB82_tc_coef | 4025 | Temp comp coeff | Temperature Compensation Coefficient: This is valid when tem-comp type is Automatic and Auto TC option is TC-coeff. The Temp Coeff is applied as percent of change of conductivity from a reference temp per deg C. Valid range is 0 to 9.99% per deg C. | R/W | Float | 8 |
var_TB82_var_ref_temp | 4026 | Variable ref temp | Variable Reference Temperature provides the ability to use use any temp as the reference temp used to determine the temp-compensated conductivity value (i.e., Primary value). | R/W | Float | 8 |
var_TB82_user_defined_tc_t1 | 404A | Temp 1 | Temp 1 of user defined temperature coefficient. First end point of the curve. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t2 | 404B | Temp 2 | Temp 2 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t3 | 404C | Temp 3 | Temp 3 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t4 | 404D | Temp 4 | Temp 4 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t5 | 404E | Temp 5 | Temp 5 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t6 | 404F | Temp 6 | Temp 6 of user defined temperature coefficient. Last end point of the curve. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k1 | 4050 | K1/Kref | Ratio of uncompensated conductivity at Temp 1 to conductivity at reference temperature (K1/Kref). First endpoint of the curve. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k2 | 4051 | K2/Kref | Ratio of uncompensated conductivity at Temp 2 to conductivity at reference temperature (K2/Kref). Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k3 | 4052 | K3/Kref | Ratio of uncompensated conductivity at Temp 3 to conductivity at reference temperature (K3/Kref). Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k4 | 4053 | K4/Kref | Ratio of uncompensated conductivity at Temp 4 to conductivity at reference temperature (K4/Kref). Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k5 | 4054 | K5/Kref | Ratio of uncompensated conductivity at Temp 5 to conductivity at reference temperature (K5/Kref). Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k6 | 4055 | K6/Kref | Ratio of uncompensated conductivity at Temp 6 to conductivity at reference temperature (K6/Kref). Last endpoint of the curve. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_conc_x1 | 4014 | x1 Cond | x1 Conductivity point of the user defined concentration curve. | R/W | Float | 8 |
var_TB82_user_defined_conc_x2 | 4016 | x2 Cond | x2 Conductivity point of the user defined concentration curve. | R/W | Float | 8 |
var_TB82_user_defined_conc_x3 | 4018 | x3 Cond | x3 Conductivity point of the user defined concentration curve. | R/W | Float | 8 |
var_TB82_user_defined_conc_x4 | 401A | x4 Cond | x4 Conductivity point of the user defined concentration curve. | R/W | Float | 8 |
var_TB82_user_defined_conc_x5 | 401C | x5 Cond | x5 Conductivity point of the user defined concentration curve. | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |